Related Experiment Videos
Evaluation of antithrombogenic thermodilution catheter
S Nagaoka1, M Mikami, Y Noishiki
1Basic Research Laboratories, Toray Industries, Inc., Kamakura, Japan.
Journal of Biomaterials Applications
|July 1, 1989
Summary
A new antithrombogenic coating, Anthron, prevents blood clots on pulmonary artery catheters. Anthron-coated catheters maintained function for over 3 days in animal studies, unlike control catheters.
Area of Science:
- Biomaterials Science
- Medical Devices
- Cardiovascular Research
Background:
- Pulmonary artery catheterization, using Swan-Ganz catheters, carries risks of thrombosis and catheter malfunction.
- Developing effective antithrombogenic coatings is crucial for improving patient safety and device longevity.
Purpose of the Study:
- To evaluate the antithrombogenic properties and functional performance of a novel heparin-releasing coating, Anthron, for thermodilution catheters.
- To compare the efficacy of Anthron-coated catheters against standard poly (vinyl chloride) catheters in preventing thrombus formation.
Main Methods:
- Development and application of an antithrombogenic coating (Anthron) onto thermodilution catheters.
- In vivo animal experiments comparing Anthron-coated catheters with control (poly (vinyl chloride)) catheters.
- Assessment of thrombus formation on catheter surfaces and surrounding blood vessels, and evaluation of catheter sensing function over time.
Main Results:
- Anthron-coated thermodilution catheters demonstrated complete freedom from thrombus formation.
- Control catheters exhibited severe thrombus formation on both the catheter surface and blood vessels.
- Anthron-coated catheters maintained excellent sensing functions for over 3 days, while control catheters showed functional deterioration due to blood constituent deposition.
Conclusions:
- The Anthron coating provides excellent antithrombogenicity and maintains catheter function, significantly reducing the risk of complications associated with pulmonary artery catheterization.
- Anthron represents a promising advancement in medical device coatings for cardiovascular applications.
- Continuous heparin release from the Anthron surface is effective in preventing thrombus formation in vivo.