A Prospective, Single Arm, Multi-site, Clinical Evaluation of a Nonradioactive Surgical Guidance Technology for the

Charles E Cox1, Scott Russell1, Vanessa Prowler2

  • 1USF Breast Health Program, University of South Florida, Tampa, FL, USA.

Abstract

Insights

The SAVI SCOUT system offers a wire-free, non-radioactive method for localizing nonpalpable breast lesions. This study confirms its high success rate in reflector placement and lesion removal, improving surgical outcomes.

Area of Science:

  • Medical Devices
  • Surgical Technology
  • Oncology

Background:

  • Nonpalpable breast lesions require accurate localization for effective surgical removal.
  • Traditional localization methods often involve wires or radioactive seeds, which can cause patient discomfort and logistical challenges.

Purpose of the Study:

  • To evaluate the SAVI SCOUT system, a micro-impulse radar technology, for breast lesion localization and surgical guidance.
  • To validate the efficacy and reliability of the SCOUT system in a multicenter trial.

Main Methods:

  • A prospective, multicenter trial involving 154 patients undergoing surgery for nonpalpable breast lesions.
  • SAVI SCOUT reflectors were placed up to 7 days pre-surgery and confirmed radiographically.
  • Surgical specimens were analyzed for reflector presence and margin status.

Main Results:

  • Successful reflector placement in 153 of 154 patients (99.4%).
  • All 154 lesions and reflectors were successfully removed during surgery.
  • For patients with cancer (n=101), 85.1% achieved clear margins on initial excision.

Conclusions:

  • The SAVI SCOUT system is a reliable and effective wire-free, non-radioactive alternative for localizing and excising nonpalpable breast lesions.
  • The system demonstrated excellent acceptance among patients, radiologists, and surgeons.