Left Ventricular Assist Device Inflow Angle and Pump Positional Change Over Time Adverse Impact on Left Ventricular
Toshinobu Kazui1, Andrew Zhang1, Jason Greenberg1
1Division of Cardiothoracic Surgery, Washington University School of Medicine, Barnes-Jewish Hospital, St. Louis, Missouri.
Pump positional change in HeartMate II left ventricular assist devices (LVADs) may contribute to dysfunction. However, changes in pump position alone do not fully explain LVAD issues in patients.
Area of Science:
- Cardiovascular Surgery
- Biomedical Engineering
- Medical Device Technology
Background:
- Investigates left ventricular assist device (LVAD) inflow cannula angulation and pump positional changes.
- Examines the incidence of thrombotic pump dysfunction in HeartMate II (HM2) patients.
Purpose of the Study:
- To analyze the relationship between LVAD inflow cannula angulation, pump position changes, and pump dysfunction.
- To identify factors contributing to HeartMate II pump dysfunction in a large patient cohort.
Main Methods:
- Retrospective review of 326 primary HM2 LVAD patients (2008-2013).
- Comparison of patients who underwent pump exchange (PE) for dysfunction, had pump dysfunction (PD) without exchange, and had normal function (NL).
- Measurement of inflow cannula (IL) angle and pump positional change from chest radiographs at multiple time points.
Main Results:
- Significant differences in IL angle were observed between PE and NL groups at all measured time points (p < 0.05).
- Sixty-seven percent of the PE group experienced pump positional change, compared to 36% in the NL group (p=0.019).
- A majority of patients with positional change or acute angulation did not develop pump dysfunction.
Conclusions:
- Pump positional change is a potential contributor to LVAD dysfunction or failure.
- Pump positional change does not solely account for all observed LVAD pump dysfunction or failure.