Counting Rate Characteristics and Image Distortion in Preclinical PET Imaging During Radiopharmaceutical Therapy
Emma Mellhammar1, Magnus Dahlbom2, Johan Axelsson3
1Division of Oncology and Pathology, Department of Clinical Sciences-Lund, Lund University, Lund, Sweden emma.mellhammar@med.lu.se.
Abstract:
PET may provide important information on the response during radiopharmaceutical therapy (RPT). Emission of radiation from the RPT radionuclide may disturb coincidence detection and impair image resolution. In this study, we tested the feasibility of performing intratherapeutic PET on 3 preclinical PET systems.
Methods:
Using 22Na point sources and phantoms filled with 18F, as well as a phantom filled with either 99mTc or 177Lu, we evaluated the coincidence counting rate and spatial resolution when both a PET and a therapeutic radionuclide were in the PET system. Because 99mTc has a suitable half-life and is easy obtainable, we used it as a substitute for a generic therapeutic radionuclide.
Results:
High activities of 99mTc deteriorated the coincidence counting rate from the 18F-filled phantom and the 22Na point source on all 3 systems. The counting rate could be corrected to a high degree on one of the systems by its dead-time correction. Spatial resolution was degraded at high 99mTc activities for all systems. On one of the systems, 177Lu increased the coincidence counting rate and slightly affected the spatial resolution. The results for high 177Lu activities were similar to those for 99mTc.
Conclusion:
Intratherapeutic imaging might be a feasible method of studying the response to RPT. However, some sensitive preclinical PET systems, unable to handle high counting rates, will have count losses and may also introduce image artifacts.
More Related Videos
09:49A Whole Body Dosimetry Protocol for Peptide-Receptor Radionuclide Therapy PRRT: 2D Planar Image and Hybrid 2D+3D SPECT/CT Image Methods
Published on: April 24, 2020
10:04Radionuclide-fluorescence Reporter Gene Imaging to Track Tumor Progression in Rodent Tumor Models
Published on: March 13, 2018
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...