Analysis of Volatile and Oxidation Sensitive Compounds Using a Cold Inlet System and Electron Impact Mass Spectrometry
Published on: September 5, 2014
Discontinuous atmospheric pressure interface for mass spectrometry using a solenoid pulse valve
Dilshadbek T Usmanov1,2, Kenzo Hiraoka1
1Clean Energy Research Center, University of Yamanashi, Takeda-4, Kofu, Yamanashi, 400-8511, Japan.
Rationale:
For the development of on-site mass spectrometry for security and safety, point-of-care analysis, etc., the gas volume introduced into the vacuum should be reduced to a minimum. To cope with this demand, a discontinuous atmospheric pressure interface using a solenoid pulse valve was developed.
Methods:
The sample gas was introduced discontinuously into the ionization cell with a volume of 0.17 cm(3) . The sampled gas in the cell was ionized by an alternating current (ac) corona discharge. The generated ions were sampled through a 0.25 mm i.d. and 12 mm long nickel capillary into the vacuum of a time-of-flight mass spectrometer.
Results:
A gas flow rate of ~25 mL/min was achieved with the 1 Hz pulse valve operation and 20 ms valve opening time. Sub-ng limits of detection for less volatile compounds such as explosives and drugs were obtained.
Conclusions:
Due to its compact size and low gas load to the vacuum, this new interface may be useful for applications in miniaturized mass spectrometry. Copyright © 2016 John Wiley & Sons, Ltd.
Related Concept Videos
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Mass Spectrometers
Gas Chromatography: Sample Injection Systems
Two primary injection methods are used...
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
High-Performance Liquid Chromatography: Instrumentation
Inductively Coupled Plasma Atomic Emission Spectroscopy: Instrumentation
There are three main types of inductively coupled plasma atomic emission spectroscopy (ICP-AES) instruments: sequential, simultaneous multichannel, and Fourier transform instruments, with the latter being less commonly used....