Related Experiment Video
Updated: Mar 17, 2026

Autonomous and Rechargeable Microneurostimulator Endoscopically Implantable into the Submucosa
Published on: September 27, 2018
Inductive Powering of Subcutaneous Stimulators: Key Parameters and Their Impact on the Design Methodology
Carmen Godfraind1, Adrien Debelle2, Laurent Lonys2
1Synergia Medical, Mont-Saint-Guibert, Belgium; Université Libre de Bruxelles, Belgium.
Abstract:
Inductive powering of implantable medical devices involves numerous factors acting on the system efficiency and safety in adversarial ways. This paper lightens up their role and identifies a procedure enabling the system design. The latter enables the problem to be decoupled into four principal steps: the frequency choice, the magnetic link optimization, the secondary circuit and then finally the primary circuit designs. The methodology has been tested for the powering system of a device requirering a power of 300mW and implanted at a distance of 15 to 30mm from the outside power source. It allowed the identification of the most critical parameters. A satisfying efficiency of 34% was reached at 21mm and tend to validate the proposed design procedure.

