Limitations and pitfalls of 99mTc-EDDA/HYNIC-TOC (Tektrotyd) scintigraphy
Ildikó Garai1, Sandor Barna, Gabor Nagy
1ScanoMed Ltd, Department of Diagnostic Imaging, University of Debrecen. garai.ildiko@scanomed.hu.
Abstract:
Tektrotyd kit was developed by Polatom company for 99mTc labeling to make an alternative tracer of somatostatin receptor scintigraphy available. Since 2005, 99mTc-EDDA/HYNIC-Tyr3-Octreotide has been used in clinical imaging and achieved high impact in management of patients with neuroendocrine tumors. Knowing the limitations and pitfalls is essential to provide ac-curate diagnosis. Therefore, the potential pitfalls associated with the use of 99mTc-EDDA/HYNIC-TOC are reviewed on the basis of own experience. Data were analyzed of 310 patients who underwent somatostatin receptor scintigraphy with 99mTc-Tektrotyd. Pitfalls during radiolabeling process or acquisition can worsen the sensitivity of SRS (somatostatin receptor scintigraphy). Recognizing physi-ological and clinical pitfalls, the diagnostic accuracy will improve.