Inter-lead correlation analysis for automated detection of cable reversals in 12/16-lead ECG
Irena Jekova1, Vessela Krasteva1, Remo Leber2
1Institute of Biophysics and Biomedical Engineering, Bulgarian Academy of Sciences, Sofia, Bulgaria.
Background And Objective:
A crucial factor for proper electrocardiogram (ECG) interpretation is the correct electrode placement in standard 12-lead ECG and extended 16-lead ECG for accurate diagnosis of acute myocardial infarctions. In the context of optimal patient care, we present and evaluate a new method for automated detection of reversals in peripheral and precordial (standard, right and posterior) leads, based on simple rules with inter-lead correlation dependencies.
Methods:
The algorithm for analysis of cable reversals relies on scoring of inter-lead correlations estimated over 4s snapshots with time-coherent data from multiple ECG leads. Peripheral cable reversals are detected by assessment of nine correlation coefficients, comparing V6 to limb leads: (I, II, III, -I, -II, -III, -aVR, -aVL, -aVF). Precordial lead reversals are detected by analysis of the ECG pattern cross-correlation progression within lead sets (V1-V6), (V4R, V3R, V3, V4), and (V4, V5, V6, V8, V9). Disturbed progression identifies the swapped leads.
Results:
A test-set, including 2239 ECGs from three independent sources-public 12-lead (PTB, CSE) and proprietary 16-lead (Basel University Hospital) databases-is used for algorithm validation, reporting specificity (Sp) and sensitivity (Se) as true negative and true positive detection of simulated lead swaps. Reversals of limb leads are detected with Se = 95.5-96.9% and 100% when right leg is involved in the reversal. Among all 15 possible pairwise reversals in standard precordial leads, adjacent lead reversals are detected with Se = 93.8% (V5-V6), 95.6% (V2-V3), 95.9% (V3-V4), 97.1% (V1-V2), and 97.8% (V4-V5), increasing to 97.8-99.8% for reversals of anatomically more distant electrodes. The pairwise reversals in the four extra precordial leads are detected with Se = 74.7% (right-sided V4R-V3R), 91.4% (posterior V8-V9), 93.7% (V4R-V9), and 97.7% (V4R-V8, V3R-V9, V3R-V8). Higher true negative rate is achieved with Sp > 99% (standard 12-lead ECG), 81.9% (V4R-V3R), 91.4% (V8-V9), and 100% (V4R-V9, V4R-V8, V3R-V9, V3R-V8), which is reasonable considering the low prevalence of lead swaps in clinical environment.
Conclusions:
Inter-lead correlation analysis is able to provide robust detection of cable reversals in standard 12-lead ECG, effectively extended to 16-lead ECG applications that have not previously been addressed.
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...