Retention behavior of isomeric polycyclic aromatic sulfur heterocycles in reversed-phase liquid chromatography
Walter B Wilson1, Lane C Sander1, Miren Lopez de Alda1
1Chemical Sciences Division, Material Measurement Laboratory, National Institute of Standards and Technology, Gaithersburg, MD 20899, United States.
Abstract:
Retention indices for 70 polycyclic aromatic sulfur heterocycles (PASHs) were determined using reversed-phase liquid chromatography (LC) on a monomeric and a polymeric C18 stationary phase. Molecular shape parameters [length, breadth, thickness (T), and length-to-breadth ratio (L/B)] were calculated for all the compounds studied. Correlations between the retention on the polymeric C18 phase and PASH geometry (L/B and T) were investigated for six specific PASH isomer groups with molecular mass (MM) 184Da, 234Da, 258Da, 284Da, 334Da, and 384Da. Similar to previous studies for polycyclic aromatic hydrocarbons (PAHs), PASH elution order on the polymeric C18 phase was generally found to follow increasing L/B values. Correlation coefficients for retention vs L/B ranged from r=0.45 (MM 184Da) to r=0.89 (MM 284Da). In the case of smaller PASHs (MM≤258Da), the location of the sulfur atom in the bay-region of the structure resulted in later than expected elution of these isomers based on L/B. In the case of the larger PASHs (MM≥284Da), nonplanarity had a significant influence on earlier than predicted elution based on L/B values.
More Related Videos
13:36Simple In-House Ultra-High Performance Capillary Column Manufacturing with the FlashPack Approach
Published on: December 4, 2021
10:14Synthesis and Purification of Iodoaziridines Involving Quantitative Selection of the Optimal Stationary Phase for Chromatography
Published on: May 16, 2014
Related Concept Videos
High-Performance Liquid Chromatography: Introduction
In HPLC, two phases play a critical role in the separation process:
Silica Gel Column Chromatography: Overview
Polar components tend to bind strongly to the silica gel, causing them to move slowly through the column. In contrast, nonpolar compounds...
Size-Exclusion Chromatography
Silica particles offer advantages such as rigidity,...
Chirality at Nitrogen, Phosphorus, and Sulfur
A consequence of chirality is the need for enantiomeric resolution. While this is theoretically possible for all...
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...