The lasso-loop, lasso-mattress and simple-cinch stitch for arthroscopic rotator cuff repair: are there biomechanical
Emmanouil Liodakis1, Antonios Dratzidis2, Manuel Kraemer3
1Trauma Department, Hannover Medical School (MHH), Carl-Neuberg-Str. 1, 30625, Hannover, Germany. liodakis.emmanouil@mh-hannover.de.
Purpose:
Various stitching techniques have been described to facilitate arthroscopic repair of rotator cuff tears. The aim of the present study was to compare the biomechanical properties of the lasso-loop, lasso-mattress and simple-cinch stitch for rotator cuff repair.
Methods:
Twelve infraspinatus tendons were harvested from sheep and split in half. The tendons were randomized into three different stitch configuration groups for biomechanical testing: lasso-loop, lasso-mattress and simple-cinch stitch. Each specimen was first cyclically loaded on a universal materials testing machine under force control from 5 to 30 N at 0.25 Hz for twenty cycles. Then, each specimen was loaded to failure under displacement control at a rate of 1 mm/s. Cyclic elongation, peak-to-peak displacement and ultimate tensile load were reported as mean ± standard error and compared using one way analysis of variance. The type of failure was recorded.
Results:
No differences in cyclic elongation (1.31 ± 0.09 mm for the simple-cinch vs. 1.49 ± 0.07 mm for the lasso-mattress vs. 1.61 ± 0.09 mm for the lasso-loop stitch, p = 0.063) or peak-to-peak displacement (0.58 ± 0.04 mm for the simple-cinch, 0.50 ± 0.03 mm for the lasso-mattress and 0.62 ± 0.06 mm for the lasso-loop stitch, p = 0.141) were seen between all tested stitch configurations. In the load-to-failure test, the simple cinch stitch (149.38 ± 11.89 N) and the lasso-mattress (149.38 ± 10.33 N) stitch demonstrated significantly higher ultimate load than the lasso-loop stitch (65.88 ± 4.75 N, p < 0.001). All stitch configurations failed with suture pull out.
Conclusions:
The lasso-mattress and the simple-cinch stitch showed similar biomechanical properties with significant higher tensile loads needed for failure than the lasso-loop stitch.