A Probing Device for Quantitatively Measuring the Mechanical Properties of Soft Tissues during Arthroscopy
Published on: May 1, 2020
Handheld Tissue Hardness Meters for Assessing the Mechanical Properties of Skeletal Muscle: A Feasibility Study
Kentaro Chino1, Hideyuki Takahashi1
1Department of Sports Science, Japan Institute of Sports Sciences, Tokyo, Japan.
Objectives:
The purpose of this study was to examine the feasibility of using handheld tissue hardness meters to assess the mechanical properties of skeletal muscle.
Methods:
This observational study included 33 healthy men (age, 22.4 ± 4.4 years) and 33 healthy women (age, 23.7 ± 4.2 years). Participants were placed in a supine position, and tissue hardness overlying the rectus femoris and the shear modulus of the muscle were measured on the right side of the body at 50% thigh length. In the same position, subcutaneous adipose tissue thickness and muscle thickness were measured using B-mode ultrasonography. To examine the associations of subcutaneous adipose tissue thickness, muscle thickness, and muscle shear modulus with tissue hardness, linear regression using a stepwise bidirectional elimination approach was performed.
Results:
Stepwise linear regression revealed that subcutaneous adipose tissue thickness (r = -0.38, P = .002) and muscle shear modulus (r = 0.27, P = .03) were significantly associated with tissue hardness.
Conclusions:
Significant associations among adipose tissue thickness, muscle shear modulus, and tissue hardness show the limitations and feasibility of handheld tissue hardness meters for assessing the mechanical properties of skeletal muscles.