Cardiac implantable electronic devices in tracheotomized muscular dystrophy patients: Safety and risks
Abdallah Fayssoil1, Arnaud Lazarus2, Karim Wahbi3
1Intensive Care Unit, GNHM Neuromuscular Center, Raymond Poincare Hospital, Garches, France; Clinical Investigation Center CIC 14.29, INSERM, Garches, France.
Background/Objectives:
Muscular dystrophies are genetic muscle disorders, in which heart involvement and chronic respiratory impairment affect survival. Cardiac conduction disturbances require implantable cardiac pacemaker. Implantable defibrillators may also be necessary to prevent cardiac sudden death. The safety and risk of cardiac electronic devices' implantation are not known in patients with muscular dystrophy. We aimed to assess the risks related to cardiac implantable electronic devices (CIED) in muscular dystrophy patients ventilated by tracheostomy.
Methods:
We reviewed all medical charts of neuromuscular patients and identified all CIED implantations of pacemakers (PM) or defibrillators (ICD) in patients ventilated using tracheostomy.
Results:
Twelve device implantations were included, performed in 9 patients (5 DMD, 1 Becker muscular dystrophy and 3 DM1). Mean age was 39.9years±13.0. All patients were wheel-chair bound and tracheotomized. Six pacemakers (PM) and 6 cardiac resynchronization (CRT) devices, including 2 defibrillators (CRT-D) were implanted. Following device implantation, two patients had a pneumothorax and one died from severe heart failure after an unsuccessful CRT implant attempt. Follow-up lasted up to 8years (mean 2.6±2.9years), during which one patient presented a PM pocket infection, requiring PM explantation and epicardial reimplantation.
Conclusion:
We found a high prevalence of early complications (16.6% pneumothorax) after CIED implantation and an acceptable long-term infectious risk (8.3%). These results highlight the feasibility of CIED implantation in tracheotomized patients with muscular dystrophies and the need for a particular caution in the management of these patients during invasive procedures. ClinicalTrials.gov (identifier: NCT02501083).
Related Concept Videos
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask...
Tracheostomy Care I: Pre-procedural Steps
Required Equipment
The equipment necessary for tracheostomy care includes: