Estimation of ambient dose equivalent distribution in the 18F-FDG administration room using Monte Carlo simulation
Shuji Nagamine1, Toshioh Fujibuchi2, Yoshiyuki Umezu3
1Division of Radiology, Department of Medical Technology, Kyushu University Hospital, 3-1-1 Maidashi, Higashi-ku, Fukuoka, 812-8582, Japan. nagamine@r-tec.med.kyushu-u.ac.jp.
Abstract:
In this study, we estimated the ambient dose equivalent rate (hereafter "dose rate") in the fluoro-2-deoxy-D-glucose (FDG) administration room in our hospital using Monte Carlo simulations, and examined the appropriate medical-personnel locations and a shielding method to reduce the dose rate during FDG injection using a lead glass shield. The line source was assumed to be the FDG feed tube and the patient a cube source. The dose rate distribution was calculated with a composite source that combines the line and cube sources. The dose rate distribution was also calculated when a lead glass shield was placed in the rear section of the lead-acrylic shield. The dose rate behind the automatic administration device decreased by 87 % with respect to that behind the lead-acrylic shield. Upon positioning a 2.8-cm-thick lead glass shield, the dose rate behind the lead-acrylic shield decreased by 67 %.
More Related Videos
Related Concept Videos
Biological Effects of Radiation
Model Approaches for Pharmacokinetic Data: Distributed Parameter Models
The distributed parameter models are specifically designed to account for variations and differences in some drug classes. This model is particularly useful for assessing regional concentrations of anticancer or...
Mechanistic Models: Compartment Models in Individual and Population Analysis
Two-Compartment Open Model: Extravascular Administration
The absorption exponent (ka) indicates the speed at which the drug...
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Dose Size and Dosing Frequency: Determination Methods