Evaluation of Two Enzyme-Linked Immunosorbent Assay Kits for Chikungunya Virus IgM Using Samples from Deceased Organ

Harry E Prince1, Michelle L Altrich2, Marek J Nowicki2

  • 1Serology Laboratory, Viracor-IBT Laboratories, Los Angeles, California, USA harry.prince@viracoribt.com.

Insights

Chikungunya virus (CHIKV) testing in organ donors is crucial. Two ELISAs were evaluated; one showed false positives, while the other indicated true CHIKV infection, highlighting the need for optimized donor screening.

Area of Science:

  • Virology
  • Immunology
  • Transplant Science

Background:

  • Increasing chikungunya virus (CHIKV) cases in returning travelers raise concerns for organ and tissue transplant safety.
  • Potential transmission of CHIKV to recipients via infected organ/tissue donors necessitates donor screening methods.
  • Evaluation of diagnostic tests for CHIKV infection in potential donors is critical for preventing transmission.

Purpose of the Study:

  • To evaluate the performance of two commercially available enzyme-linked immunosorbent assays (ELISAs) for detecting chikungunya virus (CHIKV) IgM in U.S. organ and tissue donors.
  • To assess the accuracy and reliability of these ELISAs in identifying recent CHIKV infections among donors.
  • To understand the limitations of current CHIKV IgM ELISAs and explore optimization strategies for donor screening.

Main Methods:

  • Evaluated two CHIKV IgM ELISAs (Euroimmun and InBios) using 1,000 deidentified serum/plasma specimens from organ donors (Nov 2014–Mar 2015).
  • Confirmed reactive specimens using immunofluorescence assays (IFAs) and plaque reduction neutralization testing (PRNT) at reference laboratories.
  • Employed background subtraction modification for the InBios assay to identify potential false reactivity.

Main Results:

  • The Euroimmun indirect ELISA yielded 38 falsely reactive specimens, negative by IFA for both CHIKV IgM and IgG.
  • The InBios IgM-capture ELISA identified 26 reactive samples; one specimen remained reactive after modification and showed CHIKV-specific IgG and PRNT reactivity, despite a negative IgM IFA.
  • This single InBios reactive specimen suggests true CHIKV infection, indicating potential utility but also limitations in IgM detection.

Conclusions:

  • Current CHIKV IgM ELISAs demonstrate limitations in specificity, with potential for false-positive results in organ donor screening.
  • The InBios assay, with modifications, showed promise for detecting true CHIKV infection, even with discordant IgM IFA results.
  • Understanding ELISA limitations and optimizing testing strategies are essential to prevent unnecessary organ/tissue wastage due to inaccurate CHIKV screening.