Performance of Two Different Pulse Oximeters in Neonatal Transition
Sunita Jeyapal1, Peter Prasanth Kumar Kommu1, M Manikandan2
1Department of Pediatrics, Pondicherry Institute of Medical Sciences, Ganapathichettikulam, Pondicherry, 605014, India.
Objective:
To evaluate the performance of two different pulse oximeter technologies by comparing the time taken to achieve reliable oxygen saturation readings during neonatal transition.
Methods:
This cross-sectional study was done to compare the performance of two pulse oximeter technologies - Signal Extraction Technology (SET), and Resistor Calibration (RCAL) technology on 150 inborn, term neonates at birth. Probes of both pulse oximeters were connected to the pre-ductal limb and the time taken to obtain a reliable reading was recorded. Sequential saturation values were recorded every minute till ten minutes of life. Statistical analysis was done with Mann-Whitney U test and intraclass correlation coefficient was calculated.
Results:
Pulse oximeter with SET recorded reliable readings faster than RCAL technology - Median (IQR) 32 (21-60)s vs. 45 (21.75-105)s, p = 0.021. There was a significant difference in the time to record first saturation readings of the two pulse oximeters when used in normal deliveries [RCAL vs. SET-Median (IQR) 50s (25-120.75) vs. 32.5s (21.75-58.25), p = 0.004] but no such difference was observed in Lower Segment Cesarean section (LSCS) [RCAL vs. SET Median (IQR) 35.5s (18.25-70.75) vs. 31s (20-69.75), p = 0.968]. Sequential saturation readings for every minute for the first 10 min of life were higher with SET vs. RCAL technology. The intraclass correlation coefficient between both pulse oximeters, for each minute, was poor (r < 0.80).
Conclusions:
The SET pulse oximeter picked up first saturation values faster than the RCAL technology pulse oximeters. Studies are needed to evaluate effect of this on decisions made during neonatal resuscitation.
More Related Videos
14:28Non-Invasive Monitoring of Microvascular Oxygenation and Reactive Hyperemia using Hybrid, Near-Infrared Diffuse Optical Spectroscopy for Critical Care
Published on: May 10, 2024
11:39Non-invasive Optical Measurement of Cerebral Metabolism and Hemodynamics in Infants
Published on: March 14, 2013
Related Concept Videos
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...
Guidelines For Measuring Vital Signs
Before taking a patient's vital signs, a nurse would consider and assess the patient's comfort level and ensure appropriate equipment is available.
Special considerations while measuring oxygen saturation
Ensuring accuracy in vital sign recordings while prioritizing patient comfort and minimizing anxiety is...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: