Sheathless Capillary Electrophoresis–Mass Spectrometry for Metabolic Profiling of Biological Samples
Published on: October 1, 2016
Sheathless interface to match flow rate of capillary electrophoresis with electrospray mass spectrometry using
Yue Yin1, Gongyu Li1, Yafeng Guan2
1Department of Chemistry, School of Chemistry and Materials Science, University of Science and Technology of China (USTC), Hefei, 230026, China.
Rationale:
The flow rate match has been a great challenge when coupling capillary electrophoresis (CE) with electrospray ionization mass spectrometry (ESI-MS). Conventional CE-ESI-MS interfaces used liquid sheath flow, narrowed capillary or additional pressure to meet this requirement; sacrifice of either capillary inner diameter (i.d.) or separation efficiency is often inevitable. Thus, a regular-sized capillary-based sheathless interface would be attractive for flow rate match in CE-MS.
Methods:
The regular-sized capillary-based CE-MS interface was achieved by coupling CE with induced electrospray ionization (iESI) which was stimulated by the fact that the iESI could both achieve flow rate down to 0.2 μL/min and retain ionization efficiency. The CE-iESI-MS interface was completed with an intact separation capillary, outside the outlet end of which a metal electrode was attached for the application of alternating current (ac) high voltage (HV).
Results:
The feasibility of this CE-iESI-MS interface was demonstrated through the stable total ion chromatograms obtained by continuous CE infusion of tripropylamine with regular-sized capillaries. Tripropylamine and atenolol were separated and detected successfully in phosphate buffer solution (PBS) by CE-iESI-MS using a 50 or 75 μm i.d. capillary. Furthermore, this new interface showed a better signal-to-noise (S/N) of 3 to 7 times enhancement compared with another sheathless CE-ESI-MS interface that using one high voltage for both separation and electrospray when analyzing the mixture of tripropylamine and proline in NH4 OAc buffer. In addition, the reproducibility of this interface gave satisfactory results with relative standard deviation (RSD) in retention time in the range between 1% and 3%.
Conclusions:
The novel sheathless CE-MS interface introduced here could match conventional electroosmotic flow (EOF) with electrospray which could also preserve the separation efficiency and sensitivity of CE-MS. This newly developed CE-iESI-MS interface was also demonstrated to be effective for different buffers, PBS and NH4 OAc, without any additives such as methanol and acetic acid. Hence, we believe that this sheathless CE-MS interface could be operated with other nonvolatile and volatile buffers. Copyright © 2016 John Wiley & Sons, Ltd.
More Related Videos
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
Capillary Electrophoresis: Instrumentation
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...
Electrophoresis: Overview
There...
MALDI-TOF Mass Spectrometry
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview