WHAT IS THE MAXIMUM TEMPERATURE REACHED IN ELASTOMERIC DEVICES UNDER SIMULATED OPAT CONDITIONS?
Susanna van Der Merwe1, Helen Green2
1School of Pharmacy and Biomedical Sciences, University of Portsmouth.
Archives of Disease in Childhood
|August 20, 2016
Summary
Outpatient parenteral antimicrobial therapy (OPAT) stability testing needs realistic temperatures. This study found elastomeric devices worn under clothing reach up to 28.1°C, informing safer drug administration protocols.
Area of Science:
- Pharmaceutical Sciences
- Drug Delivery Systems
- Clinical Pharmacy
Background:
- Outpatient parenteral antimicrobial therapy (OPAT) is increasingly used for cost savings in the NHS.
- There is a lack of stability data for antibiotics in elastomeric devices under real-world conditions.
- Current stability testing guidelines assume a worst-case temperature of 37°C, but evidence for this is lacking.
Purpose of the Study:
- To determine the actual temperature reached by solutions in elastomeric devices under simulated patient use.
- To provide data to inform more realistic stability testing conditions for OPAT medications.
- To investigate the influence of clothing on device temperature.
Main Methods:
- Two types of elastomeric infusion devices (Baxter Infusor LV 10 and I-Flow Homepump Eclipse® C-Series) were used.
- Miniature Type K Thermocouples monitored temperature in devices filled with sterile water.
- Devices were subjected to refrigeration followed by ambient conditions, simulating daily patient activities including wearing under clothing.
Main Results:
- The Baxter Infusor (BI) showed slower temperature changes compared to the I-Flow (IFHE) device, both during cooling and warming.
- Maximum temperatures reached when worn under clothing were 28.1°C for the BI and 23.4°C for the IFHE.
- Average maximum temperatures when worn without clothing were 23.3°C (BI) and 23.8°C (IFHE).
Conclusions:
- The study demonstrates that temperatures experienced by OPAT devices under clothing are often lower than the assumed 37°C.
- A suggested testing temperature of 30°C is proposed to account for body temperature variations and warmer ambient conditions.
- These findings support more realistic stability testing to prevent the unnecessary exclusion of drugs from OPAT.