Related Experiment Video
Updated: Mar 16, 2026

Fabrication of Three-dimensional Paper-based Microfluidic Devices for Immunoassays
Published on: March 9, 2017
Novel concept of washing for microfluidic paper-based analytical devices based on capillary force of paper substrates
Saeed Mohammadi1, Lori Shayne Alamo Busa1,2, Masatoshi Maeki3
1Graduate School of Chemical Sciences and Engineering, Hokkaido University, Kita 13 Nishi 8, Kita-ku, Sapporo, 060-8628, Japan.
Abstract:
A novel washing technique for microfluidic paper-based analytical devices (μPADs) that is based on the spontaneous capillary action of paper and eliminates unbound antigen and antibody in a sandwich immunoassay is reported. Liquids can flow through a porous medium (such as paper) in the absence of external pressure as a result of capillary action. Uniform results were achieved when washing a paper substrate in a PDMS holder which was integrated with a cartridge absorber acting as a porous medium. Our study demonstrated that applying this washing technique would allow μPADs to become the least expensive microfluidic device platform with high reproducibility and sensitivity. In a model μPAD assay that utilized this novel washing technique, C-reactive protein (CRP) was detected with a limit of detection (LOD) of 5 μg mL-1. Graphical Abstract A novel washing technique for microfluidic paper-based analytical devices (μPADs) that is based on the spontaneous capillary action of paper and eliminates unbound antigen and antibody in a sandwich immunoassay is reported.