Evaluation of newly-developed aniseikonia testing method based on space eikonometry
Kakeru Sasaki1,2, Katsuhiko Kobayashi1, Chie Usui2
1Department of Orthoptics, Teikyo University Faculty of Medical Technology, Tokyo, Japan.
Clinical & Experimental Optometry
|August 25, 2016
Summary
A new aniseikonia testing method, space eikonometry (ATSE), is more accurate than the conventional new aniseikonia tests (NAT). ATSE provides a more precise quantitative determination of aniseikonia for clinical use.
Area of Science:
- Ophthalmology
- Optometry
- Vision Science
Background:
- A novel aniseikonia testing method, space eikonometry (ATSE), has been developed.
- The conventional new aniseikonia tests (NAT) is widely used in Japan for aniseikonia assessment.
Purpose of the Study:
- To compare the accuracy of aniseikonia measurements obtained by the ATSE method against the conventional NAT.
Main Methods:
- Thirty healthy subjects with normal binocular vision were recruited.
- Aniseikonia was simulated using afocal lenses (5% and 10%) in the right eye.
- Measurements were taken three times using both ATSE and NAT, with averages used for analysis.
Main Results:
- ATSE yielded results closer to theoretical values (mean error +2%) compared to NAT (mean error -42%).
- ATSE measured 5.1% ± 0.8% and 10.1% ± 0.8% aniseikonia for 5% and 10% lenses, respectively.
- NAT measured 2.8% ± 1.3% and 6.0% ± 2.5% aniseikonia for 5% and 10% lenses, respectively.
Conclusions:
- The ATSE method demonstrates superior accuracy for quantitative aniseikonia determination.
- ATSE is recommended over NAT for precise clinical assessment of aniseikonia.
Related Concept Videos
Atomic Emission Spectroscopy: Lab
770
AES is a powerful analytical technique, especially effective when used with plasma sources, producing abundant spectra in characteristic emission lines. The Inductively Coupled Plasma (ICP), in particular, yields superior quantitative analytical data due to its high stability, low noise, low background, and minimal interferences under optimal experimental conditions. However, newer air-operated microwave sources are emerging as promising alternatives that could be more cost-effective than...
770
Electronic Distance Measuring Instruments
630
Electronic Distance Measuring Instruments (EDMs) are essential tools in modern surveying, offering precise distance measurements by emitting electromagnetic signals and calculating the time required for these signals to travel to a target and return. Two primary types of signals are used in EDMs — light waves and microwaves — each suited to specific environmental and distance requirements. Light-wave-based EDMs utilize either infrared or laser light, providing high accuracy over...