VivaSight: a new era in the evolution of tracheal tubes
Ayten Saracoglu1, Kemal T Saracoglu1
1Department of Anesthesiology and Reanimation, Istanbul Bilim University Medical School, Istanbul, Turkey.
Study Objective:
To evaluate the available data describing the use of single and double lumen VivaSight tubes.
Design:
Systematic review.
Setting:
The use of VivaSight tubes for elective surgeries including advantages, disadvantages, and possible complications.
Patients:
Systematic review of randomized controlled trials from databases including Medline, Web of Knowledge, Google Scholar, and Cochrane Central Register of Controlled Trials.
Interventions:
Comparison of VivaSight single and double-lumen tubes with conventional tubes during normal airway and expected difficult airway management. The effectiveness of the devices was also evaluated during 1-lung ventilation for patients undergoing thoracic surgery.
Measurements:
Intubation time, success rate, the requirement for fiberoptic bronchoscope, and the rate of complications.
Main Results:
Following a VivaSight double-lumen tube, a flexible bronchoscope is still needed. It is difficult to agree that VivaSight tube reduces the need or use of a bronchoscope. According to the current literature, it is unclear if there is any advantage of the VivaSight compared with using flexible bronchoscopy to direct a blocker into the correct lung. The cost may be another issue. Studies comparing VivaSight tubes with standard double lumen tubes reported faster tracheal intubation rate and higher success rate at first attempt for VivaSight. However, VivaSight tubes may cause soft tissue trauma such as bleeding, hematoma, edema, and erythema. Sore throat and dysphonia are other reported complications. Due to the outer thickness, smaller-sized double-lumen tube may be necessary. It has been reported to have the disadvantages, such as melting due to the heat of light source before insertion and sudden shutdown without warning.
Conclusions:
Problems such as overheating and melting on the distal end of the tube due to the light source and potential breakdowns of the cable should be solved by the manufacturer. This will probably require a redesign and necessitate further studies.
Related Concept Videos
Trachea
Anatomical Features:
Location: About half of the trachea is situated in the neck, anterior to the esophagus, and extends from the larynx (at the level of...
Anatomy of Respiratory System II: Lower Respiratory Tract
The Larynx
It is located between the pharynx and the trachea, acts as a passageway for air, and hosts several critical structures, such as the epiglottis, vocal cords, and glottis. The epiglottis acts as a gateway, guiding food to the...
The Bronchial Tree
The trachea, commonly known as the windpipe, is a tube that connects the larynx (voice box) to the bronchi. At a point called the carina, it bifurcates into two primary bronchi. The right primary bronchus is wider, shorter, and more vertical than the left primary...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
The Respiratory System
Anatomy of Respiratory System I: Upper Respiratory Tract
Nose and nasal cavity
The nose and nasal cavity represent the main external openings of the respiratory tract....