Related Experiment Video
Updated: Mar 15, 2026

Time-resolved Photophysical Characterization of Triplet-harvesting Organic Compounds at an Oxygen-free Environment Using an iCCD Camera
Published on: December 27, 2018
Technical brief: Pump-probe paradigm in an integrating cavity to study photodecomposition processes
Federico Gonzalez-Fernandez1, Brandi Betts-Obregon2, Andrew T Tsin2
1Departments of Ophthalmology and Pathology, University of Mississippi Medical Center, Jackson, MS; Research & Development Service, G.V. (Sonny) Montgomery Veterans Affair Medical Center, Jackson, MS.
Purpose:
Assaying photodecomposition is challenging because light must be used to initiate the photodamage and light must be used to monitor the photodecomposition. The experimental requirements are as follows: 1) During exposure of the actinic beam, continuously monitor the spectral characteristics of the sample, 2) uniformly expose the reactants to the actinic source, 3) obtain informative spectra in the presence of light scatter, and 4) achieve sufficient sensitivity for dilute reactants. Traditional spectrophotometers cannot address these issues due to sample turbidity, the inability to uniformly expose the cuvette contents to the incident beam, the inability to simultaneously perform spectral scans, and inherent low sensitivity. Here, we describe a system that meets these challenges in a practical way.
Methods:
Light access to a 8.6 ml quartz integrating sphere containing 10 µM all-trans retinol in PBS was provided by three ports at right angles allowing for the following: 1) actinic light delivery from light-emitting diodes (LEDs) firing at 100 pulses/sec, 2) entry of a separate scanning beam at 100 scans/sec (10,000 µsec scan time) via an OLIS RSM 1000 ultraviolet/visual (UV/Vis) rapid-scanning spectrophotometer (RSM), and 3) light exit to the detector photomultiplier. The RSM spectral intermediate slit was partially covered to allow for a "dark" period of 2,000 µsec when no scanning light was admitted to the cuvette. During that interval, the LED was flashed, and the photomultiplier was temporarily blocked by a perforated spinning shutter disk. The absorbance per centimeter, which is increased due to the internal reflectance of the integrating sphere compared to a standard 1 cm rectangular cuvette, was calculated according to Fry et al. (2010) Applied Optics 49:575. Retinoid photodecomposition was confirmed with high-performance liquid chromatography (HPLC).
Results:
Using the RSM to trigger the LED flash and photomultiplier shutter closure during the "dark" period allowed actinic flashes to be placed between scans. Exposure of the all-trans retinol to 366 nm flashes resulted in marked reduction in absorbance and a blue shift of the λmax. A white LED, despite its higher photon output, did not support all-trans retinol photolysis. Singular value decomposition (SVD) analysis revealed three spectral intermediates with mechanism, I -> II -> III. HPLC analysis of the reactants at the beginning and the conclusion of the light exposure confirmed the retinol photodecomposition.
Conclusions:
The highly reflecting cavity acts as a multipass cuvette that markedly increased the light path length and, thus, sensitivity. Triggering the LED during a dark period within the scan time allowed the actinic flashes to be interleafed between scans in a pump-probe paradigm. Furthermore, the entire sample was exposed to scan beam and actinic flashes, which is not possible in traditional spectrophotometers. Finally, the integrating cavity cuvette allowed use of turbid samples. SVD was useful for resolving spectral intermediates. Although the identity of the intermediates was not determined here, the ability to define molecular intermediates during photodecomposition reactions will allow future studies to isolate and identify the degradation products and determine the mechanism of light-induced retinoid degradation and that of retinoid-binding protein-mediated photoprotection.
More Related Videos
08:22Measurement of Ultrafast Vibrational Coherences in Polyatomic Radical Cations with Strong-Field Adiabatic Ionization
Published on: August 6, 2018
08:12Author Spotlight: Exploring Light-Driven Chemical Reactions and Energy-Harnessing Devices in Photochemical Research
Published on: February 16, 2024
Related Concept Videos
Double Resonance Techniques: Overview
Spin decoupling is usually achieved by...
¹³C NMR: Distortionless Enhancement by Polarization Transfer (DEPT)
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Gas Chromatography: Types of Detectors-II