Continuous Venous-Arterial Doppler Ultrasound During a Preload Challenge
Published on: January 20, 2023
Compound Doppler ultrasound signal simulation for pulsatile carotid arteries with a stenosis
Lian Gao1, Yufeng Zhang1, Yi Zhou1
1Department of Electronic Engineering, Information School, Yunnan University, Kunming, Yunnan 650091, China. E-mails: 962245641@qq.com, yfengzhang@yahoo.com, 3002380@qq.com, 1162019872@qq.com, 617674101@qq.com, ghcai@ynu.edu.cn, jhzhang@ynu.edu.cn.
Background:
The simulated Doppler blood flow signals are widely used to assess the performance of the clutter filters for removing wall components while reserving low-velocity signals scattered from physiological blood flow approaching the inner vessel-wall injured by a stenosed lesion.
Objective:
By simultaneously taking into account the natural attributes of the Doppler equipment, blood flow as well as vessel wall of pulsatile carotid arteries with a stenosis, a computer simulation method is presented to produce the compound Doppler ultrasound blood flow signals.
Methods:
The in-phase and quadrature (I/Q) axial as well as radial blood flow signals are simulated by superposing a series of cosine functions regulated by the spectrograms estimated from the axial and radial velocity profiles firstly obtained through the solution of the incompressible Navier-Stokes equations, respectively. Meanwhile, the I/Q Doppler signals echoed from pulsatile near (anterior) and far (posterior) walls are reproduced based on their radial movements during a cardiac cycle. Ultimately, those confirmed quadrature signals are summed to generate the compound Doppler signals including the contribution from both blood flow and stenosed vessel-wall.
Results:
The compound Doppler ultrasound signals echoed from both axial and radial blood flows as well as vessel walls with obstruction grades of 0% (normal arteries), 10% and 25% are simulated respectively. The real signals from the left carotid artery with an approximately 10% stenosis degree are also collected for further assessing the believability of simulated versions.
Conclusions:
The simulated and clinical tests demonstrate that the proposed computer simulation method can produce compound Doppler signals with confirmed qualitative and quantitative characteristics resembled with the clinical versions, which could be used as an theoretical data source for evaluating the performance of the signal separation between pulsatile blood flows and vessel walls with mild stenosed-lesions.
More Related Videos
09:36A Magnetic Resonance Imaging-based Computational Protocol for Analysis of Plaque Morphology and Hemodynamics in Patients with Carotid Artery Stenosis
Published on: August 12, 2025
09:24O-Ring Aortic Banding Versus Traditional Transverse Aortic Constriction for Modeling Pressure Overload-Induced Cardiac Hypertrophy
Published on: October 6, 2022
Related Concept Videos
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Equipments Used To Measure Blood Pressure
This invasive approach involves cannulating a peripheral artery. During each cardiac contraction, pressure generates mechanical motion within the catheter, transmitted through rigid, fluid-filled tubing to a transducer. This transducer converts mechanical motion into electrical signals displayed as waveforms on a monitor. An automatic flushing system prevents blood backflow. Due to the potential risk of unexpected arterial blood loss, this method is primarily used in intensive...
Pulse
The pulse serves as a clinical...