Feasibility of spectral shaping for detection and quantification of coronary calcifications in ultra-low dose CT
Marleen Vonder1,2, Gert Jan Pelgrim1,2, Sèvrin E M Huijsse1
1Department of Radiology, University of Groningen, University Medical Center Groningen, Hanzeplein 1, EB44, 9713 GZ, Groningen, The Netherlands.
Objectives:
To evaluate detectability and quantification of coronary calcifications for CT with a tin filter for spectral shaping.
Methods:
Phantom inserts with 100 small and 9 large calcifications, and a moving artificial artery with 3 calcifications (speed 0-30 mm/s) were placed in a thorax phantom simulating different patient sizes. The phantom was scanned in high-pitch spiral mode at 100 kVp with tin filter (Sn100 kVp), and at a reference of 120 kVp, with electrocardiographic (ECG) gating. Detectability and quantification of calcifications were analyzed for standard (130 HU) and adapted thresholds.
Results:
Sn100 kVp yielded lower detectability of calcifications (9 % versus 12 %, p = 0.027) and lower Agatston scores (p < 0.008), irrespective of calcification, patient size and speed. Volume scores of the moving calcifications for Sn100 kVp at speed 10-30 mm/s were lower (p < 0.001), while mass scores were similar (p = 0.131). For Sn100 kVp with adapted threshold of 117 HU, detectability (p = 1.000) and Agatston score (p > 0.206) were similar to 120 kVp. Spectral shaping resulted in median dose reduction of 62.3 % (range 59.0-73.4 %).
Conclusions:
Coronary calcium scanning with spectral shaping yields lower detectability of calcifications and lower Agatston scores compared to 120 kVp scanning, for which a HU threshold correction should be developed.
Key Points:
• Sn100kVp yields lower detectability and lower Agatston scores compared to 120kVp • Adapted HU threshold for Sn100kVp provides Agatston scores comparable to 120kVp • Sn100 kVp considerably reduces dose in calcium scoring versus 120 kVp.
More Related Videos
06:57Author Spotlight: Advancing Cardiovascular Imaging - Introducing the Spatially Weighted Calcium Score for Early Disease Detection
Published on: September 22, 2023
08:02Author Spotlight: Enhanced Quantification of Cardiovascular Calcification Progression for Longitudinal Micro PET/CT Studies in Small Research Animals
Published on: November 15, 2024
Related Concept Videos
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...