Relationship between pull-out strength and oscillation angle in bipolar cups: an in vitro study
1Department of Orthopaedic Surgery, Faculty of Medicine, Oita University, Yufu City, Oita, Japan.
Purpose:
To compare the past and present bipolar hip arthroplasty (BHA) models in terms of balance between pull-out strength and oscillation angle (OA).
Methods:
The pull-out strength and OA of 8 BHA models were compared: UPF-II, IBC, and Tandem XLPE (Smith & Nephew); Ringloc x (Biomet); J-FX (DePuy); Bipolar (Nakashima Medical); Multipolar (Zimmer); and Centrax (Stryker).
Results:
Respectively for the UPF-II, IBC, Tandem, Ringloc, J-FX, Nakashima Bipolar, Multipolar, and Centrax, the mean pull-out strength was 2219 N, 3303 N, 1503 N, 951 N, 1453 N, 1856 N, 1536 N, and 753 N, whereas the mean OA was 54.2°, 53.8°, 64.0°, 73.2°, 63.0°, 65.4°, 55.6°, and 75.4°. The OA was lower in the integrated types. For pull-out strength of the locking mechanism, the integrated type (IBC and Nakashima) was stronger than the metal or polyethylene ring-lock type (all others). The pull-out strength and OA were negatively correlated (r= -0.881, p=0.007), and the balance between the 2 varied for different models.
Conclusion:
There is a trade-off between the pullout strength and OA; optimal balance between the 2 should be based on each patient's need.
More Related Videos
09:21Experiments on Ultrasonic Lubrication Using a Piezoelectrically-assisted Tribometer and Optical Profilometer
Published on: September 28, 2015
08:59Modeling and Experimental Analysis of the Single-Shaft Coaxial Motor-Pump Assembly in Electrohydrostatic Actuators
Published on: June 13, 2022
Related Concept Videos
Angle of Twist: Problem Solving
Torsion of Noncircular Members
Bearings: Problem Solving
Unsymmetric Bending - Angle of Neutral Axis
When a bending moment is applied at an angle θ concerning the vertical axis of a symmetrical member, it can be resolved into components along the member's principal...
Thin-Walled Hollow Shafts
Deformation in a Circular Shaft