Francesco Giannini1, Matteo Montorfano2, Vittorio Romano2
1Interventional Cardiology Unit, San Raffaele Hospital, Milan, Italy; EMO-GVM Centro Cuore Columbus, Milan, Italy.
Even with advanced transcatheter aortic valve implantation (TAVI) devices, valve embolization can occur. Severe left ventricular hypertrophy and septal anatomy are critical factors influencing TAVI valve stability.
Area of Science:
- Cardiology
- Interventional Cardiology
- Medical Device Technology
Background:
- Transcatheter aortic valve implantation (TAVI) has revolutionized aortic valve replacement.
- First-generation TAVI valves were associated with rare but serious complications like valve embolization.
- Second-generation TAVI valves were developed with enhanced retrievability and repositionability to mitigate risks.
Observation:
- This report details the first documented case of embolization involving a second-generation, fully retrievable Lotus valve.
- The complication occurred in a patient with severe left ventricular hypertrophy and a sigmoid septum.
- Initial implantation of a 23mm Lotus valve at a high position may have been compromised by septal bulging, leading to distal embolization.
Findings:
- A subsequent implantation with a 25mm Lotus valve, utilizing a lower height and oversizing for increased radial strength, achieved greater stability.
- The "watermelon seeding" phenomenon, where the septum dislodges the valve, was observed.
- This case underscores the potential for valve embolization even with newer TAVI technologies.
Implications:
- Meticulous pre-TAVI assessment of subvalvular geometry, especially severe hypertrophy, is crucial for device selection and implantation strategy.
- Consideration of self-expanding prostheses or lower implantation heights with oversized valves may enhance stability.
- Higher implantation of the Lotus valve in hypertrophic patients may increase the risk of embolization and necessitate careful planning.
More Related Videos
06:59Improved Registration of 3D CT Angiography with X-ray Fluoroscopy for Image Fusion During Transcatheter Aortic Valve Implantation
Published on: June 3, 2018
14:14Standardized Technique of Aortic Valve Re-implantation for Valve-sparing Aortic Root Replacement
Published on: December 11, 2017
Related Concept Videos
Mitral Stenosis III: Medical Management
Venous Thrombosis III: Interprofessional Care
Aortic Regurgitation III: Medical Management