Related Experiment Video
Updated: Mar 15, 2026

High-Resolution Cardiac Positron Emission Tomography/Computed Tomography for Small Animals
Published on: December 16, 2022
Standardization of 99mTechnetium pyrophosphate imaging methodology to diagnose TTR cardiac amyloidosis
Sabahat Bokhari1, Rachelle Morgenstern2, Richard Weinberg2
1Department of Medicine and Division of Cardiology, Columbia University Medical Center, 622 West 168th street, PH10-203E, New York, NY, 10032, USA. sb605@cumc.columbia.edu.
Background:
Technetium pyrophosphate (99mTc-PYP) imaging to diagnose transthyretin cardiac amyloidosis (ATTR-CA) has been increasingly utilized. The objective of this study is to provide a standardized 99mTc-PYP imaging protocol to diagnose ATTR-CA.
Methods:
104 scans from 45 subjects with biopsy-proven ATTR-CA or light-chain cardiac amyloidosis (AL) were assessed. Multiple scans were obtained using different counts (750 vs 2000 K), times to acquisition (1 vs 2 to 4 hours), processing matrix (256 vs 128), and 99mTc-PYP dose. Image quality and extracardiac activity was assessed. Quantitative methods using heart-to-contralateral ratios (H/CL) and a visual semiquantitative scale were used to diagnose ATTR-CA.19 The correlation between H/CL ratios and reproducibility of semiquantitative visual scores, acquired using various imaging parameters, were evaluated.
Results:
All imaging parameters had good to excellent image quality. 750 vs 2000 K counts, 1 hour acquisition and 256 matrix, had lower extracardiac activity (P = .00018). 10 mCi of 99mTc-PYP v. higher doses showed excellent image quality and less extracardiac activity (P = .0015). Correlation of H/CL ratios was strong (r ≥ 0.92) and reproducibility of semiquantitative visual scores was high (Kappa = 95%).
Conclusion:
An imaging protocol using 750 K counts, 10 mCi of 99mTc-PYP, and a 256 matrix was chosen as the standardized imaging protocol since it provided the shortest overall study time (1 vs 2 to 4 hours) and lowest radiation exposure (3 vs 8 to 10 mSv).
More Related Videos
11:17A Microscopic 2,3,5-Triphenyltetrazolium Chloride Assay for Accurate and Reliable Analysis of Myocardial Injury
Published on: November 28, 2025
09:34Dynamic Imaging of Chimeric Antigen Receptor T Cells with [18F]Tetrafluoroborate Positron Emission Tomography/Computed Tomography
Published on: February 17, 2022
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System V: CT
Positron Emission Tomography
One of the main requirements of a PET scan is a positron-emitting radioisotope, which is produced in a cyclotron and then attached to a substance used by the part of the body...