Inappropriate fixation of an endotracheal tube causing cuff malfunction resulting in difficult extubation
Deb Sanjay Nag1, Devi Prasad Samaddar1
1Department of Anaesthesiology & Critical Care, Tata Main Hospital, Jamshedpur, Jharkhand, India.
Abstract:
We discuss a case of difficult extubation, due to inadequate deflation of the tracheal tube cuff, despite collapse of the pilot balloon, on its aspiration. This was caused by inadvertent kinking of the pilot balloon tubing due to inappropriate tape fixation of the endotracheal tube.
Related Concept Videos
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals....
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Endotracheal Intubation II: Nursing Management
1. Nursing Care of Patients Before Intubation
Before the endotracheal intubation procedure, nurses play an essential role in ensuring the process goes smoothly. The nurses must be familiar with intubation...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques