Abstract:
Thiosulfate measurement is crucial to diagnosis of hydrogen sulfide (H2S) poisoning in forensic toxicology. Although GC-MS method is currently regarded as a standard thiosulfate measurement, it requires complicated sample preparation prior to analysis. This study presents a simple, rapid, and highly sensitive method for the quantitative analysis of serum thiosulfate by using liquid chromatography-tandem mass spectrometry (LC-MS/MS). This method is based on selected reaction monitoring and has high sensitivity with a lower quantification limit of 0.5μM. Precision and accuracy of this method meet the basic requirements for quantitative analysis (intra- and inter-day tests have a relative standard deviation of ⩽10.4%; range of analytical recovery is 94.3-102.6%). On the measurements of serum thiosulfate by our developed method, a thiosulfate concentration as 57.5μM was detected clearly in the H2S poisoning case comparing to the non poisoning case in which only a trace amount of thiosulfate was observed.