Reinfusate Heparin Concentrations Produced by Two Autotransfusion Systems.
Willem Frederik Buys1, Mathilde Buys1, Andrew Ian Levin1
1Department of Anesthesiology and Critical Care, Faculty of Health Sciences, University of Stellenbosch and Tygerberg Academic Hospital, Cape Town, South Africa.
Cell saver reinfusate heparin levels were assessed in cardiac surgery patients. The Sorin (Dideco) Electa device largely met the American Association of Blood Banks threshold, unlike the Sorin Xtra device.
Area of Science:
- Cardiovascular Surgery
- Transfusion Medicine
- Biomedical Engineering
Background:
- Cell saver reinfusate ideally has low heparin concentrations, below the American Association of Blood Banks' (AABB) 0.5 IU/mL threshold.
- Uncertainty exists regarding cell saver heparin elimination rates and reinfusate heparin concentrations.
Purpose of the Study:
- To investigate reinfusate heparin concentrations from cell saver devices used in a university teaching hospital.
- To determine if cell saver reinfusates meet clinically insignificant heparin concentration thresholds.
Main Methods:
- Two prospective, observational studies were conducted at a university teaching hospital.
- Heparin concentrations in reinfusates from Sorin (Dideco) Electa and Sorin Xtra cell savers were measured using a modified anti-Xa chromogenic assay in patients undergoing on-pump cardiac surgery.
- 32 patients were enrolled for the Electa study and 31 for the Xtra study.
Main Results:
- The Sorin Xtra device had reinfusate heparin concentrations exceeding 0.5 IU/mL in 56% of cases.
- The Sorin (Dideco) Electa device showed heparin concentrations above 0.5 IU/mL in only 9.4% of cases (3 out of 32).
- 90.6% of Sorin (Dideco) Electa reinfusates met the AABB heparin threshold.
Conclusions:
- The Sorin (Dideco) Electa device demonstrated significantly lower reinfusate heparin concentrations compared to the Sorin Xtra device.
- Accurate measurement of cell saver reinfusate heparin requires a modified chromogenic assay.
- Periodic quality control of cell saver reinfusate composition is recommended to ensure patient safety.
Related Concept Videos
Anticoagulant Drugs: Low-Molecular-Weight Heparins
Determination of Multiple Dosing Parameters: Steady-State, Minimum and Maximum Concentrations
Drug Accumulation During Multiple Dosing: Intermittent IV Infusions
Blood Transfusion and Agglutination
History
The history of blood transfusion dates back to the 17th century, when early attempts were made in animals. In 1818 James Blundell, a British doctor, performed the first successful human blood transfusion. Later in 1900, Karl...
One-Compartment Model: IV Infusion
The one-compartment model for IV infusion uses mathematical equations to describe the rate of change in drug quantity in the body. At steady-state or infusion equilibrium, the drug input...
Extracorporeal Removal of Drugs: Hemoperfusion and Hemofiltration