Single-Incision Mini-Slings: Obturator Complex Pull-Out-Force Measurements.
Florian Lenz1, Sara Doll, Christof Sohn
1Department of Obstetrics and Gynecology, St. Marien- and St. Annastiftskrankenhaus, Ludwigshafen, Germany.
Gynecologic and Obstetric Investigation
|September 7, 2016
Summary
The Ajust mini-sling demonstrated a stronger pull-out force from the obturator complex compared to the MiniArc system. Both systems, however, offer sufficient stability for treating stress urinary incontinence (SUI).
Area of Science:
- Urogynecology
- Pelvic Floor Surgery
- Biomechanical Analysis
Background:
- Stress urinary incontinence (SUI) affects many women, with mini-sling procedures offering a minimally invasive treatment option.
- The stability of mini-sling anchors within pelvic anatomy is crucial for successful SUI treatment.
- Understanding the biomechanical properties of different mini-sling systems is essential for surgical decision-making.
Purpose of the Study:
- To compare the pull-out force of two mini-sling systems (MiniArc and Ajust) from the obturator membrane and obturator internus muscle.
- To evaluate the biomechanical stability of these systems in female cadaveric models.
- To determine which system offers superior anchorage for stress urinary incontinence treatment.
Main Methods:
- A prospective study utilizing 8 female cadaver pelvises.
- Pull-out force measurements were conducted using a force measurement device (PCE-FM50).
- Hooks from MiniArc and Ajust mini-sling kits were tested for their anchorage strength in the obturator membrane and obturator internus muscle.
Main Results:
- The Ajust mini-sling exhibited a significantly higher mean pull-out force (2,561.75 ± 638.8 g) from the complete obturator complex compared to the MiniArc (1,780.69 ± 442.3 g) (p = 0.025).
- No significant differences in pull-out force were observed between the two systems when anchored solely to the obturator membrane.
- These findings indicate differential biomechanical performance based on the specific anatomical structure targeted.
Conclusions:
- The Ajust mini-sling hook demonstrates more stable attachment to the obturator complex than the MiniArc hook.
- Both tested mini-sling systems provide sufficient anchorage to the evaluated anatomical structures for SUI treatment.
- The biomechanical data suggest potential differences in long-term efficacy related to anchor stability.