A new method to measure oxygenator oxygen transfer performance during cardiopulmonary bypass: clinical testing using
Carole Hamilton1,2, Denise Marin1, Frank Weinbrenner1
11 Cardiac Surgery, Schoen Klinik Vogtareuth, Vogtareuth, Germany.
Background:
There is no acceptable method of testing oxygen transfer performance in membrane oxygenators quickly and easily during cardiopulmonary bypass. Pre-clinical testing of oxygenators is performed under controlled situations in the laboratory, correlating oxygen transfer to blood flow using 100% oxygen. This laboratory method cannot be used clinically as oxygen transfer values vary significantly at each blood flow and the FiO2 is not kept at 1. Therefore, a formula was developed which corrects the existing FiO2 to attain a PaO2 of 150 mmHg: the corrected FiO2 at 150 mmHg. In graph form, this corrected FiO2 (x-axis) is correlated to the patient's oxygen consumption levels (y-axis), which determines the membrane oxygenator oxygen transfer performance.
Methods:
Blood gas and hemodynamic parameters taken during cardiopulmonary bypass using the Medtronic Fusion were used to calculate the oxygen consumption (inlet conditions to the oxygenator) and the corrected FiO2 for a PaO2 of 150 mmHg. Validation of the formula "FiO2-PaO2/(Pb-pH2O)+0.21" was carried out by plotting the calculated values on a graph using PaO2 values between 145 to 155 mmHg and then, using the corrected FiO2 for PaO2s outside of this range.
Results:
All trend-lines correlated significantly to confirm that the Medtronic Fusion had an extrapolated oxygen transfer of 419 milliliters O2/min at an FiO2 of 1 to achieve a PaO2 of 150 mmHg.
Conclusions:
Use of the corrected FiO2 correlated to the oxygen transfer conditions of the membrane oxygenator can easily be used on a routine basis, providing valuable information clinically. When used by the manufacturer under laboratory conditions, further clinically relevant data is provided in terms of FiO2 and resultant PaO2s instead of the present limitations using blood flow. In this way, a clinically justifiable method has been developed to finally establish a standard in testing membrane oxygenator performance.
More Related Videos
06:22Author Spotlight: Advancing Cardiovascular Research — Tailored Langendorff Perfusion Techniques for Improved Experimental Outcomes
Published on: June 14, 2024
09:47Author Spotlight: Utilizing Next-Generation Polymerized Human Hemoglobin for Improved Donor Lung Evaluation and Preservation in Rats
Published on: June 14, 2024
Related Concept Videos
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Assessment of Diffusion and Perfusion
The Role of Diffusion in Respiration
Diffusion is the process by which molecules move from an area of higher concentration to an area of lower concentration. In the respiratory system, this...
Oxygen Transport in the Blood
Pulse Oximetry
Purpose
Average SpO2 values are greater than 95%. If the readings fall below 90%, it indicates that...