Thin film based semi-active resonant marker design for low profile interventional cardiovascular MRI devices
Engin Baysoy1,2, Dursun Korel Yildirim3, Cagla Ozsoy3
1Institute of Biomedical Engineering, Bogazici University, Istanbul, Turkey. enginbaysoy@gmail.com.
Objectives:
A new microfabrication method to produce low profile radio frequency (RF) resonant markers on catheter shafts was developed. A semi-active RF resonant marker incorporating a solenoid and a plate capacitor was constructed on the distal shaft of a 5 Fr guiding catheter. The resulting device can be used for interventional cardiovascular MRI procedures.
Materials And Methods:
Unlike current semi-active device visualization techniques that require rigid and bulky analog circuit components (capacitor and solenoid), we fabricated a low profile RF resonant marker directly on guiding the catheter surface by thin film metal deposition and electroplating processes using a modified physical vapor deposition system.
Results:
The increase of the overall device profile thickness caused by the semi-active RF resonant marker (130 µm thick) was lowered by a factor of 4.6 compared with using the thinnest commercial non-magnetic and rigid circuit components (600 µm thick). Moreover, adequate visibility performance of the RF resonant marker in different orientations and overall RF safety were confirmed through in vitro experiments under MRI successfully.
Conclusion:
The developed RF resonant marker on a clinical grade 5 Fr guiding catheter will enable several interventional congenital heart disease treatment procedures under MRI.
More Related Videos
07:13Author Spotlight: An Efficient and Robust Software for Automated Fusion of Multiple Preclinical Imaging Modalities
Published on: October 27, 2023
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
Related Concept Videos
Imaging Studies for Cardiovascular System IV: CMRI
Magnetic Resonance Imaging