A Comprehensive Performance Evaluation of Five Blood Glucose Systems in the Hypo-, Eu-, and Hyperglycemic Range
Eric Zijlstra1, Lutz Heinemann2, Annelie Fischer2
1Profil, Neuss, Germany eric.zijlstra@profil.com.
Background:
The objective was to evaluate the performance (in terms of accuracy, precision, and trueness) of 5 CE-certified and commercially available blood glucose (BG) systems (meters plus test strips) using an innovative clinical-experimental study design with a 3-step glucose clamp approach and frequent capillary BG measurements.
Methods:
Sixteen subjects with type 1 diabetes participated in this open label, single center trial. BG was clamped at 3 levels for 60 minutes each: 60-100-200 mg/dL. Medical staff performed regular finger pricks (up to 10 per BG level) to obtain capillary blood samples for paired BG measurements with the 5 BG systems and a laboratory method as comparison.
Results:
Three BG systems displayed significantly lower mean absolute relative deviations (MARD) (ACCU-Chek® Aviva Nano [5.4%], BGStar® [5.1%], iBGStar® [5.3%]) than 2 others (FreeStyle InsuLinx® [7.7%], OneTouch Verio®IQ [10.3%]). The measurement precision of all BG systems was comparable, but relative bias was also lower for the 3 systems with lower MARD (ACCU-Chek [1.3%], BGStar [-0.9%], iBGStar [1.0%]) compared with the 2 others (FreeStyle [-7.2%], OneTouch [8.9%]).
Conclusions:
This 3 range glucose clamp approach enables a systematic performance evaluation of BG systems under controlled and reproducible conditions. The random error of the tested BG systems was comparable, but some showed a lower systematic error than others. These BG systems allow an accurate glucose measurement at low, normal and high BG levels.
More Related Videos
07:35Author Spotlight: Investigating the Blood Glucose Homeostasis in Murine Brain Using a Cost-Effective Hyperglycemic And Hypoglycemic Clamp Technique
Published on: January 26, 2024
08:01A Method for Manipulating Blood Glucose and Measuring Resulting Changes in Cognitive Accessibility of Target Stimuli
Published on: August 12, 2016
Related Concept Videos
Hypoglycemia and Glucagon
Diabetes: Symptoms, Diagnosis, and Complications
Feedback Loops
Glucose Homeostasis: Regulation of Blood Glucose
During fasting, when blood glucose levels are low, the pancreas secretes glucagon. it...
Hormones Regulating Blood Glucose
In addition to accelerating glucose uptake and utilization, insulin has...
Errors occurring during blood pressure monitoring
Several factors...