Related Experiment Video
Updated: Mar 15, 2026

Author Spotlight: Advancing Human Brain Modulation – Optimized Protocols for Transcranial Ultrasound Stimulation Experiments
Published on: June 28, 2024
Measured acoustic intensities for clinical diagnostic ultrasound transducers and correlation with thermal index
K Retz1, S Kotopoulis2,3, T Kiserud4,5
1Department of Obstetrics and Gynecology, Helse Stavanger, Stavanger, Norway.
Objectives:
To investigate if the thermal index for bone (TIB) displayed on screen is an adequate predictor for the derated spatial-peak temporal-average (ISPTA.3 ) and spatial-peak pulse-average (ISPPA.3 ) acoustic intensities in a selection of clinical diagnostic ultrasound machines and transducers.
Methods:
We calibrated five clinical diagnostic ultrasound scanners and 10 transducers, using two-dimensional grayscale, color Doppler and pulsed-wave Doppler, both close to and far from the transducer, with a TIB between 0.1 and 4.0, recording 103 unique measurements. Acoustic measurements were performed in a bespoke three-axis computer-controlled scanning tank, using a 200-μm-diameter calibrated needle hydrophone.
Results:
There was significant but poor correlation between the acoustic intensities and the on-screen TIB. At a TIB of 0.1, the ISPTA.3 range was 0.51-50.49 mW/cm2 and the ISPPA.3 range was 0.01-207.29 W/cm2 . At a TIB of 1.1, the ISPTA.3 range was 19.02-309.44 mW/cm2 and the ISPPA.3 range was 3.87-51.89 W/cm2 .
Conclusions:
TIB is a poor predictor for ISPTA.3 and ISPPA.3 and for the potential bioeffects of clinical diagnostic ultrasound scanners. Copyright © 2016 ISUOG. Published by John Wiley & Sons Ltd.
More Related Videos
Related Concept Videos
Ultrasonography
During an ultrasonography procedure, a handheld device called...
Ultrasound II: Endoscopic Ultrasound and FibroScan
Endoscopic Ultrasound (EUS):
Sound Intensity Level
The human ear can perceive an extensive range of sound intensity, necessitating the use of the logarithmic scale to define a physical quantity—the intensity level. It is a ratio of two intensities and...
Ultrasound I: Abdominal Ultrasonography
Abdominal ultrasonography, commonly known as abdominal ultrasound, is a vital, non-invasive medical imaging technique widely used in healthcare.
Procedure:
This diagnostic tool allows the clinician to visually inspect internal structures within the abdomen, including vital organs such as the liver, gallbladder, pancreas, kidneys, and spleen.
The abdominal ultrasound process begins with applying a special gel to the patient's skin over the abdomen. This gel enhances the...
Imaging Studies II: Ultrasonography
Intensity and Pressure of Sound Waves
Unlike the time average of a sinusoidal term, which is zero since it is positive...