Optimal flexible laryngeal mask airway size in children weighing 10 to 20 kg
1Anesthesiologist, Department of Anesthesiology, Shanghai Eye, Ear, Nose, and Throat Hospital, Fudan University, Shanghai, China.
Insights
Choosing the correct flexible laryngeal mask airway (FLMA) size is crucial for pediatric patients. While both size 2 and 2.5 FLMA showed similar oropharyngeal leak pressure (OLP) in smaller children, size 2.5 is preferable for children weighing 16-20 kg to ensure adequate ventilation.
Area of Science:
- Anesthesiology
- Pediatric Airway Management
- Medical Device Evaluation
Background:
- Securing a pediatric airway during surgery requires careful device selection.
- Flexible laryngeal mask airways (FLMA) are commonly used, but optimal sizing for effective ventilation is critical.
- Oropharyngeal leak pressure (OLP) is a key indicator of airway seal efficacy.
Purpose of the Study:
- To evaluate the impact of flexible laryngeal mask airway (FLMA) size on oropharyngeal leak pressure (OLP) in pediatric patients.
- To compare OLP and ventilation adequacy between size 2 and size 2.5 FLMA in children undergoing elective ophthalmic surgery.
- To identify optimal FLMA sizing based on patient weight for effective airway management.
Main Methods:
- Prospective, randomized study involving 120 children undergoing elective ophthalmic surgery.
- Children were randomized to receive either a size 2 or size 2.5 FLMA at a recommended intracuff pressure of 40 cmH2O.
- Primary outcome was OLP; secondary outcomes included ventilation sufficiency, insertion success, and adverse events.
Main Results:
- Median OLP was comparable between size 2 and 2.5 FLMA (18 cmH2O) in the overall cohort.
- The size 2 FLMA group exhibited a higher incidence of low OLP (<10 cmH2O) and insufficient ventilation (13.3% vs. 0%, P=0.006).
- In children weighing 16-20 kg, the size 2.5 FLMA demonstrated significantly lower rates of low OLP and insufficient ventilation (27% vs. 0%, P=0.0046).
Conclusions:
- FLMA size 2 and 2.5 provide similar OLP in children weighing 10-15.9 kg at 40 cmH2O intracuff pressure.
- For pediatric patients weighing 16-20 kg, a size 2.5 FLMA is recommended to reduce the incidence of low OLP and ensure adequate ventilation.
- Appropriate FLMA sizing based on weight is essential for optimizing pediatric airway management and ventilation.
Abstract:
This prospective, randomised study was conducted to assess the effect of flexible laryngeal mask airway (FLMA) size on oropharyngeal leak pressure (OLP) in children at the recommended intracuff pressure. A total of 120 children undergoing elective ophthalmic surgery were randomly assigned to the size 2 FLMA group or size 2.5 FLMA group. The primary measurement was OLP at an intracuff pressure of 40 cmH2O. Secondary outcomes included the incidence of OLP <10 cmH2O, insufficient ventilation, gastric insufflation, insertion time, successful first-attempt insertion rate, fibreoptic view grade and pharyngolaryngeal adverse events. The median OLP was comparable for the size 2 and size 2.5 FLMA (18 cmH2O versus 18 cmH2O, P=0.38). However, the size 2 FLMA group had a higher incidence of OLP <10 cmH2O and insufficient ventilation (13.3% versus 0, P=0.006). In subgroup analyses based on weight, the size 2.5 FLMA had a lower occurrence of OLP <10 cmH2O and insufficient ventilation (27% versus 0, P=0.0046) in children 16-20 kg. We conclude that at a 40 cmH2O intracuff pressure, the OLP with the size 2 and size 2.5 FLMA was similar in children weighing 10-15.9 kg. However, in children weighing 16-20 kg, size 2 devices had a higher incidence of low OLP and insufficient ventilation, so a 2.5 FLMA may be preferable in this subgroup.
Related Concept Videos
Cardiopulmonary Resuscitation II: ACLS Airway Management
Oxygen Delivering System I: Nasal Cannula and Face Mask
Nasal Cannula
A nasal cannula is a lightweight tube split at one end into two prongs and placed in the nostrils. It is typically used to deliver low to medium levels of oxygen.
Suggested flow rate: The suggested flow rate for a nasal cannula typically ranges between 1 and 6 L/min.
Oxygen percentage setting:...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Suctioning the Nasopharyngeal Airway
Equipment Required
Assessment of Airway, Skin Color, and Use of Accessory Muscles
Introduction
The initial evaluation of a patient's respiratory system...