Related Experiment Video
Updated: Mar 15, 2026

Employing the Forced Oscillation Technique for the Assessment of Respiratory Mechanics in Adults
Published on: February 9, 2022
Early Diagnosis of Respiratory Abnormalities in Asbestos-Exposed Workers by the Forced Oscillation Technique
Paula Morisco de Sá1, Hermano Albuquerque Castro2, Agnaldo José Lopes3
1Biomedical Instrumentation Laboratory, Institute of Biology and Faculty of Engineering and BioVasc Research Laboratory, Institute of Biology, State University of Rio de Janeiro, Rio de Janeiro, Brazil.
Background:
The current reference test for the detection of respiratory abnormalities in asbestos-exposed workers is spirometry. However, spirometry has several shortcomings that greatly affect the efficacy of current asbestos control programs. The forced oscillation technique (FOT) represents the current state-of-the-art technique in the assessment of lung function. This method provides a detailed analysis of respiratory resistance and reactance at different oscillatory frequencies during tidal breathing. Here, we evaluate the FOT as an alternative method to standard spirometry for the early detection and quantification of respiratory abnormalities in asbestos-exposed workers.
Methodology/Principal Findings:
Seventy-two subjects were analyzed. The control group was composed of 33 subjects with a normal spirometric exam who had no history of smoking or pulmonary disease. Thirty-nine subjects exposed to asbestos were also studied, including 32 volunteers in radiological category 0/0 and 7 volunteers with radiological categories of 0/1 or 1/1. FOT data were interpreted using classical parameters as well as integer (InOr) and fractional-order (FrOr) modeling. The diagnostic accuracy was evaluated by investigating the area under the receiver operating characteristic curve (AUC). Exposed workers presented increased obstruction (resistance p<0.001) and a reduced compliance (p<0.001), with a predominance of obstructive changes. The FOT parameter changes were correlated with the standard pulmonary function analysis methods (R = -0.52, p<0.001). Early respiratory abnormalities were identified with a high diagnostic accuracy (AUC = 0.987) using parameters obtained from the FrOr modeling. This accuracy was significantly better than those obtained with classical (p<0.001) and InOr (p<0.001) model parameters.
Conclusions:
The FOT improved our knowledge about the biomechanical abnormalities in workers exposed to asbestos. Additionally, a high diagnostic accuracy in the diagnosis of early respiratory abnormalities in asbestos-exposed workers was obtained. This makes the FOT particularly useful as a screening tool in the context of asbestos control and elimination. Moreover, it can facilitate epidemiological research and the longitudinal follow-up of asbestos exposure and asbestos-related diseases.
Related Concept Videos
Respiratory System Abnormal Finding I: Inspection and Percussion
Inspection Findings
During an inspection, several findings may suggest the presence of respiratory distress or disease. Pursed-lip breathing, where exhalation is slowed by...
Respiratory System Abnormal Finding II: Palpation and Auscultation
Palpation Findings
During a respiratory assessment, palpation can reveal several vital abnormalities:
Assessment of Respiration
Subjective Assessment: Nurses interview the patient to gather information directly during the subjective assessment. It includes questions about the individual's medical history, medications, and symptoms, focusing on past respiratory conditions like...
Chronic Obstructive Pulmonary Disease-IV: Assessement and Diagnostic Studies
Medical History
Physical Assessment of the Respiratory Tract II: Inspection
Chest Configuration
The chest configuration...
Physical Assessment of the Respiratory Tract IV: Auscultation
Breath Sounds
Breath sounds are categorized into vesicular, bronchovesicular, and bronchial.