Application of Light-cured Dental Adhesive Resin for Mounting Electrodes or Microdialysis Probes in Chronic Experiments
Published on: July 30, 2007
Light output from six battery operated dental curing lights
Carlos Alberto Kenji Shimokawa1, Míriam Lacalle Turbino2, Jessie Eudora Harlow3
1University of São Paulo, School of Dentistry, Restorative Dentistry, Avenida Professor Lineu Prestes, 2227, 05508-000, São Paulo, São Paulo, Brazil; Dalhousie University, Faculty of Dentistry, Dental Clinical Sciences, 5981 University Avenue, B3H 4R2, Halifax, Nova Scotia, Canada.
Dental light curing units (LCUs) show significant variations in light output and beam quality. Performance differences were noted after repeated use, highlighting the need for standardized reporting by manufacturers and researchers.
Area of Science:
- Dental Materials Science
- Photopolymerization Technology
- Biomedical Engineering
Background:
- Light Curing Units (LCUs) are essential dental devices for polymerizing dental resins.
- Visual assessment of LCU performance is unreliable due to high light intensity.
- Variability in LCU light output can impact clinical outcomes.
Purpose of the Study:
- To evaluate and compare the light output characteristics of six contemporary dental LCUs.
- To assess the consistency of LCU performance after repeated light exposures.
- To analyze differences in radiant power, irradiance, emission spectrum, and beam profile.
Main Methods:
- Radiant power and irradiance were measured using a laboratory-grade power meter.
- Emission spectra were analyzed with a fiber-optic spectrometer.
- Light beam profiles were assessed for homogeneity.
- Measurements were taken with a fully charged battery, after 50, and 100 exposures.
Main Results:
- Significant differences in power and irradiance were found between LCUs (p<0.01).
- Optilight Prime and Slim Blast showed significant output reduction after 50 and 100 exposures.
- Bluephase G2 exhibited significant reduction only after 100 exposures.
- Bluephase G2 and Translux 2Wave had dual-peak emission spectra; Elipar Deep Cure-S and Bluephase G2 had homogeneous beam profiles.
Conclusions:
- Contemporary dental LCUs exhibit substantial variations in light output and beam characteristics.
- LCU performance can degrade significantly with repeated use.
- Standardized reporting of LCU light output data by manufacturers and researchers is recommended.