Right atrial lead fixation type and lead position are associated with significant variation in complications
Chance M Witt1, Charles J Lenz1, Henry H Shih2
1Division of Cardiovascular Diseases, Mayo Clinic, 200 1st St SW, Rochester, MN, 55905, USA.
Purpose:
Optimal atrial pacemaker lead position and fixation mechanism have not been determined with regard to effect on complications. We aimed to determine the association between atrial lead-related complications and varying atrial lead tip positions and lead fixation mechanisms.
Methods:
All patients who underwent dual-chamber pacemaker implant between 2004 and 2014 were retrospectively reviewed for atrial lead tip position and fixation type. Lead-related complications were assessed by electronic medical record review. Complication rates were compared at 1 year by chi-square analysis and at 5 years using a Kaplan-Meier analysis.
Results:
During the study period, 3451 patients (mean age 73.9, 53.4 % male) underwent dual-chamber pacemaker placement. Active fixation leads were associated with a higher incidence of pericardial effusion (81 (2.9 %) vs. 6 (1.0 %), p = 0.005) and pericardiocentesis (46 (1.6 %) vs. 2 (0.3 %), p = 0.01) at 1 year compared to passive fixation leads. There was no difference in overall complication rates by fixation type (161 (5.7 %) vs. 29 (4.6 %), p = 0.26). Low atrial septal lead tip position was associated with a higher rate of lead dislodgement (10 (15.2 %)) compared to appendage (46 (1.6 %)), free wall (10 (2.1 %)), or high atrial septal (2 (4.7 %)) positions (p < 0.001). This difference was also reflected in a significantly increased need for lead revision and overall complications. A multivariate analysis which included potential confounders confirmed the association of active fixation leads with an increased rate of perforation-related complications (p = 0.03) and septal lead location with increased rates of dislodgement (p < 0.001).
Conclusions:
Active compared to passive lead fixation increases the risk for pericardial effusion requiring pericardiocentesis. There is a clear association between low atrial septal lead position and lead dislodgement requiring lead revision.
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias
Cardiac Catheterization III: Left Heart Catheterization
Cardiac Catheterization II: Right Heart Catheterization