Related Experiment Videos
Simple formulae for the calculation of RHO, T1 and T2 from a properly designed diagnostic NMR experiment
1University Hospital Utrecht, The Netherlands.
Magnetic Resonance Imaging
|May 1, 1989
Abstract:
The calculation of rho, T1 and T2 images from a minimal set of source images is considerably simplified if proper values for the sequence parameters are chosen. For the spin-echo technique a setup is suggested - a ratio of three between the effective repetition time of the T2- and the T1-weighted acquisitions - which not only produces the desired simplifications but which also is entirely compatible with normal diagnostic practice and close to optimum for two-point measurements.