Early Detection of Infusion Set Failure During Insulin Pump Therapy in Type 1 Diabetes
Marzia Cescon1, Daniel J DeSalvo2, Trang T Ly3
1Department Chemical Engineering & Institute for Collaborative Biotechnologies, University of California, Santa Barbara, Santa Barbara, CA, USA.
Background:
Insulin infusion set failure resulting in prolonged hyperglycemia or diabetic ketoacidosis can occur with pump therapy in type 1 diabetes. Set failures are frequently characterized by variable and unpredictable patterns of increasing glucose values despite increased insulin infusion. Early detection may minimize the risk of prolonged hyperglycemia, an important consideration for automated insulin delivery and closed-loop applications.
Methods:
A novel algorithm designed to alert the patient to the onset of infusion set failure was developed based upon continuous glucose sensor values and insulin delivered from an insulin pump. The method was calibrated on 12 weeks of infusion set wear without failures recorded by 4 patients in ambulatory conditions and prospectively validated on 18 weeks of infusion set wear with and without failures belonging to 9 other subjects in ambulatory conditions.
Results:
The algorithm, evaluated retrospectively, identified a failure 2.52 ± 1.91 days ahead of the actual event as recorded by the clinical team, corresponding to 50% sensitivity, 66% specificity and 55% accuracy. If set failure alarms had been activated in real time, the average time >180 mg/dl would be reduced from 82.7 ± 40.9 hours/week/subject (without alarm) to 58.8 ± 31.1 hours/week/subject (with alarm), corresponding to a potential 29% reduction in time spent >180mg/dl.
Conclusion:
The proposed method for early detection of infusion set failure based on glucose sensor and insulin data demonstrated favorable results on retrospective data and may be implemented as an additional safeguard in a future fully automated closed-loop system.
Related Concept Videos
Insulin: Dosing Regimen and Adverse Effects
The basal dose constitutes about 40%-50% of the total daily dose, with the rest as premeal insulin. The mealtime insulin dose should mirror...
Diabetes Mellitus: Overview and Type I Subtype
Type 1 diabetes is an autoimmune disease in which the immune system mistakenly attacks and destroys the insulin-producing beta cells in the pancreas. As a result, the body is unable to produce sufficient insulin, and individuals with...
Insulin Formulations: Types and Delivery
Short-acting insulins are divided into...
Diabetes Mellitus: Type 2 and Gestational
Diabetes: Management and Pharmacotherapy
Insulin remains the cornerstone of treatment for most patients with type 1 and many...
Diabetes: Symptoms, Diagnosis, and Complications