Background:
With recent advances in sensor and computer technologies, the ability to monitor peripheral pulse activity is no longer limited to the laboratory and clinic. Now inexpensive sensors, which interface with smartphones or other computer-based devices, are expanding into the consumer market. When appropriate algorithms are applied, these new technologies enable ambulatory monitoring of dynamic physiological responses outside the clinic in a variety of applications including monitoring fatigue, health, workload, fitness, and rehabilitation. Several of these applications rely upon measures derived from peripheral pulse waves measured via contact or non-contact photoplethysmography (PPG). As technologies move from contact to non-contact PPG, there are new challenges. The technology necessary to estimate average heart rate over a few seconds from a noncontact PPG is available. However, a technology to precisely measure instantaneous heat rate (IHR) from non-contact sensors, on a beat-to-beat basis, is more challenging.
Objectives:
The objective of this paper is to develop an algorithm with the ability to accurately monitor IHR from peripheral pulse waves, which provides an opportunity to measure the neural regulation of the heart from the beat-to-beat heart rate pattern (i.e., heart rate variability).
Methods:
The adaptive harmonic model is applied to model the contact or non-contact PPG signals, and a new methodology, the Synchrosqueezing Transform (SST), is applied to extract IHR. The body sway rhythm inherited in the non-contact PPG signal is modeled and handled by the notion of wave-shape function.
Results:
The SST optimizes the extraction of IHR from the PPG signals and the technique functions well even during periods of poor signal to noise. We contrast the contact and non-contact indices of PPG derived heart rate with a criterion electrocardiogram (ECG). ECG and PPG signals were monitored in 21 healthy subjects performing tasks with different physical demands. The root mean square error of IHR estimated by SST is significantly better than commonly applied methods such as autoregressive (AR) method. In the walking situation, while AR method fails, SST still provides a reasonably good result.
Conclusions:
The SST processed PPG data provided an accurate estimate of the ECG derived IHR and consistently performed better than commonly applied methods such as autoregressive method.
Related Concept Videos
Correlation between ECG and Cardiac Cycle
A cardiac action potential originates in the SA node and spreads throughout the atria and the AV node in approximately 0.03 seconds. This results in the P wave in an ECG and triggers atrial contraction. The action potential is then briefly slowed at the AV node, allowing the atria to contract and fill the ventricles with blood before...
Pulse rhythm
Conversely, an irregular pulse pattern is termed dysrhythmia, stemming from disruptions in cardiac...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
Factors Influencing Heart Rate
Let us explore the significant factors affecting heart rate, including age, body temperature, posture, acute pain, chemical influences,...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment: