Objective:
The aim of this prospective randomized controlled study was to compare 2 techniques for radiation dose reduction in non-contrast-enhanced pediatric chest computed tomography (CT): low peak kilovoltage imaging at 70 kVp and spectral beam shaping at 100 kVp using a dedicated tin filter (100-kVp Sn).
Materials And Methods:
All chest CT examinations were performed on a third-generation dual-source CT system (SOMATOM Force; Siemens Healthineers, Germany). Fifty children (mean age, 6.8 ± 5.1 years) were examined using the 100-kVp Sn protocol, whereas 25 children received the 70-kVp protocol (mean age, 5.7 ± 5.2 years; 2:1 randomization scheme). Radiation metrics and organ doses were compared between acquisition techniques using commercially available radiation dose analysis software (Radimetrics Inc, Bayer AG, Toronto, Ontario, Canada). Objective image quality, expressed by signal-to-noise ratio and subjective image quality based on a 4-point scale (1, best; 4, worst image quality), were compared.
Results:
Volume CT dose index and size-specific dose estimate were significantly lower in the 100-kVp Sn group compared with the 70-kVp group (0.19 ± 0.12 mGy vs 0.81 ± 0.70 mGy and 0.34 ± 0.13 mGy vs 1.48 ± 1.11 mGy; P < 0.0001 for both). Accordingly, mean effective dose was significantly lower for the 100-kVp Sn examinations (0.21 ± 0.10 mSv) compared with the 70-kVp examinations (0.83 ± 0.49 mSv; P < 0.0001). Calculated organ doses were also significantly lower using the 100-kVp Sn protocol when compared with the 70-kVp protocol; for example, breast dose was reduced by a factor of 4.3. Signal-to-noise ratio was slightly superior for 70-kVp images while lung image quality of the 100-kVp Sn protocol was preferred in subjective analysis (P = 0.0004).
Conclusions:
Pediatric chest CT performed at 100 kVp with an additional tin filter for spectral shaping significantly reduces radiation dose when compared with low peak kilovoltage imaging at 70 kVp and therefore should be preferred in non-contrast-enhanced pediatric chest CT examinations, particularly (given the improved subjective image quality) when the main focus is evaluation of the lung parenchyma.
More Related Videos
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation II: MRI and Ventilation Perfusion Scan
Magnetic Resonance Imaging (MRI) and Ventilation Perfusion Scans are two radiological investigations that offer detailed diagnostic images of the body, particularly lung structures.
MRI
MRI uses magnetic fields and radiofrequency signals to distinguish between normal and abnormal tissues. This technology provides a more detailed diagnostic image than CT scans, enabling it to characterize pulmonary nodules, stage bronchogenic carcinoma, and evaluate inflammatory activity in...