Human factors analysis of the CardioQuick Patch®: A novel engineering solution to the problem of electrode
Raymond R Bond1, Dewar D Finlay1, James McLaughlin1
1Ulster University, Northern Ireland, UK.
Introduction:
The CardioQuick Patch® (CQP) has been developed to assist operators in accurately positioning precordial electrodes during 12-lead electrocardiogram (ECG) acquisition. This study describes the CQP design and assesses the device in comparison to conventional electrode application.
Methods:
Twenty ECG technicians were recruited and a total of 60 ECG acquisitions were performed on the same patient model over four phases: (1) all participants applied single electrodes to the patient; (2) all participants were then re-trained on electrode placement and on how to use the CQP; (3) participants were randomly divided into two groups, the standard group applied single electrodes and the CQP group used the CQP; (4) after a one day interval, the same participants returned to carry out the same procedure on the same patient (measuring intra-practitioner variability). Accuracy was measured with reference to pre-marked correct locations using ultra violet ink. NASA-TLK was used to measure cognitive workload and the Systematic Usability Scale (SUS) was used to quantify the usability of the CQP.
Results:
There was a large difference between the minimum time taken to complete each approach (CQP=38.58s vs. 65.96s). The standard group exhibited significant levels of electrode placement error (V1=25.35mm±29.33, V2=18.1mm±24.49, V3=38.65mm±15.57, V4=37.73mm±12.14, V5=35.75mm±15.61, V6=44.15mm±14.32). The CQP group had statistically greater accuracy when placing five of the six electrodes (V1=6.68mm±8.53 [p<0.001], V2=8.8mm±9.64 [p=0.122], V3=6.83mm±8.99 [p<0.001], V4=14.90mm±11.76 [p<0.001], V5=8.63mm±10.70 [p<0.001], V6=18.13mm±14.37 [p<0.001]). There was less intra-practitioner variability when using the CQP on the same patient model. NASA TLX revealed that the CQP did increase the cognitive workload (CQP group=16.51%±8.11 vs. 12.22%±8.07 [p=0.251]). The CQP also achieved a high SUS score of 91±7.28.
Conclusion:
The CQP significantly improved the reproducibility and accuracy of placing precordial electrodes V1, V3-V6 with little additional cognitive effort, and with a high degree of usability.
More Related Videos
10:17Real-Time Cardiac Mapping with a Noninvasive Imageless Electrocardiographic Imaging System
Published on: April 11, 2025
06:57Ablation of Ischemic Ventricular Tachycardia Using a Multipolar Catheter and 3-dimensional Mapping System for High-density Electro-anatomical Reconstruction
Published on: January 31, 2019
Related Concept Videos
Electrocardiogram Fundamentals
An electrocardiogram (ECG) is a diagnostic tool for identifying cardiac conditions such as arrhythmias, conduction abnormalities, and myocardial ischemia.
Definition
An electrocardiogram (ECG) visualizes the heart's electrical activity by tracing the electrical movement associated with each heartbeat on a graph or monitor. As the heart beats, an electrical wave passes through it, correlating with the cardiac cycle events.
Parts of an ECG
An ECG utilizes electrodes on the skin...
Electrocardiogram
Three major waveforms are present in a typical ECG recording: the P wave, the QRS complex, and...
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias