Erroneous HbA1c results in a patient with elevated HbC and HbF
Joy Adekanmbi1, Trefor Higgins2, Karina Rodriguez-Capote2
1Department of Laboratory Medicine and Pathology, University of Alberta, Canada.
Background:
HbA1c is used in the diagnosis and monitoring of diabetes mellitus (DM). Interference from hemoglobin variants is a well-described phenomenon, particularly with HPLC-based methods. While immunoassays may generate more reliable HbA1c results in the presence of some variants, these methods are susceptible to negative interference from high concentrations of HbF. We report a case where an accurate HbA1c result could not be obtained by any available method due to the presence of a compound hemoglobinopathy.
Methods:
HbA1c was measured by HPLC, immunoassay, and capillary electrophoresis. Hemoglobinopathy investigation consisted of a CBC, hemoglobin fractionation by HPLC and electrophoresis, and molecular analysis.
Results:
HbA1c analysis by HPLC and capillary electrophoresis gave no result. Analysis by immunoassay yielded HbA1c results of 5.9% (Siemens DCA 2000+) and 5.1% (Roche Integra), which were inconsistent with other markers of glycemic control. Hemoglobinopathy investigation showed HbC with the hereditary persistence of fetal hemoglobin-2 Ghana deletion.
Conclusion:
Reliable HbA1c results may be unobtainable in the presence of some hemoglobinopathies. HPLC and capillary electrophoresis alerted the laboratory to the presence of an unusual hemoglobinopathy. Immunoassays generated falsely low results without warning, which could lead to missed diagnoses and under treatment of patients with DM.
More Related Videos
07:24A Precision Medicine Tool for Measurement and Monitoring of Hemoglobin S in Sickle Cell Disease Patients Receiving Transfusion Therapy
13:21Chemically-blocked Antibody Microarray for Multiplexed High-throughput Profiling of Specific Protein Glycosylation in Complex Samples
Published on: May 4, 2012
Related Concept Videos
Errors In Hypothesis Tests
Errors occurring during blood pressure monitoring
Several factors...
Rh Blood Group
Diabetes: Symptoms, Diagnosis, and Complications
Hypoglycemia and Glucagon
Blood Typing
Antigens are protein molecules that reside on the surface of red blood cells (RBCs). The ABO and Rh blood typing systems target...