Multifocal Electroretinograms
Published on: December 4, 2011
Variability of the multifocal electroretinogram based on the type and position of the electrode
Ángel García-García1, Francisco José Muñoz-Negrete2, Gema Rebolleda2
1Clínica Eurocanarias Oftalmológica, León y Castillo, 211, 35004, Las Palmas de Gran Canaria, Spain. angarciagarcia@gmail.com.
Purpose:
To evaluate and compare the wave amplitude of multifocal electroretinogram (mfERG) measurements and discomfort from Dawson-Trick-Litzkow (DTL) electrode located on the cornea (cDTL) and on the conjunctival fornix (fDTL) and ERG-jet contact lens electrode (CL).
Methods:
Thirty-six patients and 18 healthy volunteers were evaluated with three different methods (cDTL, fDTL and CL). Wave amplitude, number of artifacts, number of electrode dislocations and level of discomfort obtained with each electrode were compared. These variables were also compared between patients and healthy volunteers (control group). Comparisons between wave amplitudes were made using intraclass correlation coefficient (ICC).
Results:
Responses by the three tested electrode were better correlated (ICC) in patient (0.616-0.756) than in control groups (0.342-0.679). CL provided the highest wave amplitude in both groups (p < 0.005), but it was associated with higher discomfort (p < 0.001) and the highest rate of dislocations (72 and 100 % in control and patients, respectively). Looking at the differences obtained by each electrode between both groups, CL seems to be able to differentiate patient from control in the ring 1 of the mfERG. By contrast, fDTL gave the lowest wave amplitude in both groups, but it had the advantage to, apparently, discriminate between patients and control group in rings 1 and 2. cDTL produced more artifacts than the other electrodes in both groups (p < 0.001).
Conclusions:
Although wave amplitude measurements with different electrodes were correlated, they cannot be directly compared, so it is mandatory to create an appropriate normative database with each electrode. Despite providing the lowest amplitudes, fDTL seems to offer the best features to perform mfERG regarding discomfort, number of artifacts and diagnostic capability.
More Related Videos
10:30Simultaneous Recording of Electroretinography and Visual Evoked Potentials in Anesthetized Rats
Published on: July 1, 2016
07:52Multiscale Investigations of Cortical Processing by Integrating Laminar Polytrodes and Optogenetics with Micro Electrocorticography in Rodents
Published on: May 23, 2025
Related Concept Videos
Electrodes: Overview
There are two main types of electrodes in electrochemical cells. The first type, known as the working or indicator electrode, has a potential that is sensitive to the analyte's concentration and reacts to changes in...
Potentiometry: Types of Electrodes
The Standard Hydrogen Electrode (SHE) is a widely used reference electrode that maintains zero potential across all temperatures. However, its need for a continuous hydrogen gas supply renders it impractical for everyday use.
An alternative to SHE is the Saturated Calomel Electrode (SCE). This electrode features an...
Standard Electrode Potentials