How Should Results of Nonsurgical Subcutaneous Fat Removal Be Assessed? Accuracy of B-Mode Ultrasound
Fritz E Barton1,2, Phillip B Dauwe1,2, Tara Stone1,2
1Dallas, Texas.
Background:
Nonsurgical fat reduction has become extremely popular among patients; however, a reliable method of measuring its efficacy has not been established.
Methods:
Ultrasound measurement of human female abdominal subcutaneous fat thickness was carried out on five volunteers. Forty-seven measurements were performed using a GE Venue 40 diagnostic ultrasound device with a 12-MHz transducer. Transducer pressure measurements were recorded simultaneously according to the protocol described by Toomey et al..
Results:
Reproducible measurements of abdomen subcutaneous fat could be consistently achieved with a margin of error (95 percent CI) of ±0.558 mm.
Conclusions:
Using a protocol with a transducer pressure less than 1 N (Toomey protocol) allows accurate and reliable measurement of subcutaneous fat. The authors further conclude that such a protocol is practically reproducible in the clinical setting and should be the standard for evaluating the results of nonsurgical fat removal, particularly in the abdomen.
Clinical Question/Level Of Evidence:
Diagnostic, IV.