Technical Note: Out-of-field dose measurement at near surface with plastic scintillator detector
Alexandra Bourgouin1, Nicolas Varfalvy, Louis Archambault
1Université Laval. alex_8919@hotmail.com.
Abstract:
Out-of-field dose depends on multiple factors, making peripheral dosimetry com-plex. Only a few dosimeters have the required features for measuring peripheral dose. Plastic scintillator dosimeters (PSDs) offer numerous dosimetric advantages as required for out-of-field dosimetry. The purpose of this study is to determine the potential of using PSD as a surface peripheral dosimeter. Measurements were performed with a parallel-plate ion chamber, a small volume ion chamber, and with a PSD. Lateral-dose measurements (LDM) at 0.5 cm depth and depth-dose curve (PDD) were made and compared to the dose calculation provided by a treatment planning system (TPS). This study shows that a PSD can measure a dose as low as 0.51 ± 0.17 cGy for photon beam and 0.58 ± 0.20 cGy for electron beam with a difference of 0.2 and 0.1 cGy compared to a parallel-plate ion chamber. This study demonstrates the potential of using PSD as an out-of-field dosimeter since measure-ments with PSD avoid averaging over a too-large depth, at 1 mm diameter, and can make precise measurement at very low dose. Also, electronic equilibrium is easier to reach with PSD due to its small sensitive volume and its water equivalence.
More Related Videos
11:27Studying Soft-matter and Biological Systems over a Wide Length-scale from Nanometer and Micrometer Sizes at the Small-angle Neutron Diffractometer KWS-2
Published on: December 8, 2016
10:24Neutron Radiography and Computed Tomography of Biological Systems at the Oak Ridge National Laboratory's High Flux Isotope Reactor
Published on: May 7, 2021