Non-fluoroscopic Catheter Tracking for Fluoroscopy Reduction in Interventional Electrophysiology
Published on: May 26, 2015
Safety and efficacy of applying a low-dose radiation fluoroscopy protocol in device implantations
Philipp Attanasio1, Melika Mirdamadi1, Jean-Yves Wielandts2,3
1Department of Internal Medicine and Cardiology, Charité - Universitätsmedizin Berlin, Campus Virchow-Klinikum, Augustenburger Platz 1, Berlin 13353, Germany.
Aims:
For cardiac implantable electronic device (CIED) implantations, visualization of lead placement is necessary and fluoroscopy remains by far the most commonly used technique. With simple changes in the X-ray system settings, total radiation dose can be reduced significantly. The purpose of this study was to assess the safety and efficacy of various CIED implantations performed after implementation of a new dose reduction protocol (DRP).
Methods And Results:
We conducted a retrospective chart review of 584 patients undergoing CIED implantation or revision in our hospital. Of these patients, 280 (48%) underwent the implantation prior to and 304 (52%) after the DRP introduction. The DRP included various changes for optimized image processing and exposure system settings to enable dose reduction, as well as a reduced frame rates (4 FPS for fluoroscopy and 7.5 FPS for cinematographic images). Of the 584 patients, 53 (9.1%) had a one-chamber pacemaker, 232 (39.7%) a two-chamber pacemaker, 133 (22.8%) a one-chamber ICD, 35 (6.0%) a two-chamber ICD, 82 (14.0%) a CRT (de novo) implantation, and 49 (8.3%) had an upgrade to a CRT device. DRP was associated with a 64% reduction of the dose-area product (1372 ± 2659 vs. 3792 ± 5025 cGcm2, P < 0.001), while fluoroscopy duration (13 ± 15 vs. 13 ± 15 min) and procedural duration (93 ± 52 vs. 92 ± 52 min.) did not significantly increase. Complication rates did not differ significantly between the two groups.
Conclusion:
The DRP proved to effectively reduce radiation dose for all types of CIED implantations. Fluoroscopy time, total procedure time, and the number of complications did not increase after introducing the DRP.
Related Concept Videos
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies for Cardiovascular System V: CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Radiological Investigation I: X-ray and CT
Imaging Studies VII: Vascular Imaging