Initial clinical results with the ThermoCool® SmartTouch® Surround Flow catheter
Hanney Gonna1,2, Giulia Domenichini1, Zia Zuberi1,3
1Cardiology Clinical Academic Group, St. George's University Hospitals NHS Foundation Trust, St. George's, University of London, London, UK.
The new SmartTouch Surround Flow (STSF) catheter offers improved safety and efficacy for cardiac ablation compared to the older SF catheter. This advanced catheter technology resulted in fewer complications and a better overall outcome for patients undergoing arrhythmia treatment.
Area of Science:
- Electrophysiology
- Cardiovascular Medicine
- Medical Device Technology
Background:
- The Biosense Webster ThermoCool® SmartTouch® Surround Flow (STSF) catheter, introduced in May 2015, replaced the ThermoCool® SF catheter.
- The STSF catheter features Surround Flow (SF) technology for enhanced cooling and integrated force sensing for precise tissue contact quantification.
Purpose of the Study:
- To compare the safety and efficacy of the novel STSF catheter against the previously used SF catheter.
- To evaluate the performance of the STSF catheter in a prospective cohort of 100 ablation procedures.
Main Methods:
- Prospective data collection for 100 STSF catheter procedures.
- Matched case-control analysis comparing STSF procedures with 100 SF catheter procedures from a database of 654.
- Matching criteria included procedure type, operator experience, patient age, and gender.
Main Results:
- Shorter fluoroscopy duration in the STSF group (mean 25.8 min vs. 30.0 min, P=0.03).
- No complications occurred with the STSF catheter, versus two complications (pericardial effusion, permanent pacing) in the SF group.
- The composite endpoint of procedure failure or acute complication was significantly lower in the STSF group (2 vs. 8, P=0.05).
- Complete procedural success was high in both groups (98% STSF vs. 94% SF, P=0.15).
Conclusions:
- The STSF catheter is a safe and effective tool for treating various cardiac arrhythmias.
- The STSF catheter demonstrates a trend towards an improved safety-efficacy profile compared to the SF catheter.
Related Concept Videos
Urodynamic Studies: Uroflowmetry
Cardiac Catheterization IV: Nursing Management
Assessing Body Temperature - Rectal
Follow these steps for rectal temperature assessment:
Step 1: Perform hand hygiene and don clean gloves to prevent cross-infection.
Step 2: Position the patient in a side-lying position to better visualize the rectal...