Do CAD/CAM dentures really release less monomer than conventional dentures?
Patricia-Anca Steinmassl1, Verena Wiedemair2, Christian Huck2
1University Hospital for Dental Prosthetics and Restorative Dentistry, Medical University of Innsbruck, MZA, Anichstr. 35, A-6020, Innsbruck, Austria.
Clinical Oral Investigations
|October 6, 2016
Summary
Computer-aided design/computer-aided manufacturing (CAD/CAM) dentures offer advantages like lower weight and volume. However, this study found no significant reduction in methacrylate monomer release compared to conventional dentures.
Area of Science:
- Dental prosthetics
- Materials science
Background:
- Computer-aided design/computer-aided manufacturing (CAD/CAM) dentures are increasingly used.
- It is hypothesized that CAD/CAM dentures possess superior material properties, including reduced methacrylate monomer release, compared to conventional dentures.
Purpose of the Study:
- To investigate the hypothesis that CAD/CAM dentures exhibit lower methacrylate monomer release than conventionally fabricated dentures.
- To compare physical properties such as volume, weight, density, and surface area of CAD/CAM and conventional dentures.
Main Methods:
- Four different CAD/CAM systems and conventional heat-polymerised dentures were used.
- Denture weight, volume, density, and surface area were measured.
- Methacrylate monomer release was quantified using high-performance liquid chromatography after 7 days of water storage.
Main Results:
- CAD/CAM dentures from Whole You Nexteeth and Wieland Digital Dentures showed significantly lower volume and weight than conventional dentures.
- Baltic Denture System and Whole You Nexteeth exhibited significantly increased density.
- No significant difference in methacrylate monomer release was observed between CAD/CAM and conventional dentures.
Conclusions:
- While CAD/CAM dentures offer benefits like reduced weight and volume, potentially enhancing patient comfort and mechanical properties, they did not demonstrate significantly lower methacrylate monomer release than conventional dentures.
- The study did not verify the hypothesis regarding reduced monomer release for CAD/CAM dentures compared to conventional ones.
- Further research may explore differences with other denture base materials like autopolymerising resins.